![Maico MB 11 BERAphone Operation Manual Download Page 56](http://html1.mh-extra.com/html/maico/mb-11-beraphone/mb-11-beraphone_operation-manual_3713323056.webp)
Operation Manual MB 11 BERAphone
®
/Classic
8100496 Rev. 6
54
23/10/2019
Test Results
5.5.5.1 Test Result PASS
During the test, a small segment of a line is drawn after each 1 second of data collection.
The line displays as blue for a left ear test and red for a right ear test (MB 11 Classic only:
Magenta for binaural test). During the test, the indication line for pass criteria in the diagram
continues to move upward on the graph until the green area is reached. Then 100 % of the
pass criteria are fulfilled and the test was passed successfully. The result
PASS
is shown in
the green area (see Figure 39).
Figure 39
5.5.5.2 Test Result REFER
If the pass criterion is not reached after a specified time, the result
REFER
displays in
the lower right corner (Figure 40). If the setting
Allow reduced test time
is enabled,
a
REFER
will be recorded at 120 seconds in case the progress line has never reached
60 % toward the pass criteria. Otherwise a
REFER
will be recorded at 180 seconds of
test time.
Figure 40
If you cannot exclude poor test conditions as a reason for the
REFER
you should
consider performing a retest when the conditions have improved. Follow your hearing
screening pro
gram’s protocol for handling patients with a
REFER
outcome.