Operating Instruction
ERO•SCAN
®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
66
8106556-5 02/15
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The ERO•SCAN
®
is intended for use in the electromagnetic environment specified below. The customer or
the user of the ERO•SCAN
®
should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance
Electromagnetic
Environment-Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material the
relative humidity should
be greater than 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality
should be that of a typical
commercial, hospital, or
residential environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality
should be that of a typical
commercial, hospital, or
residential environment.
Voltage Dips, Short
Interruptions and
Voltage Variations on
Power Supply Lines
IEC 61000-4-11
<5 %
U
T
(>95 % dip in
U
T) for
0.5 cycle
40 %
U
T
(60% dip in
U
T) for 5
cycles
70 %
U
T
(30% dip in
U
T) for 25
cycles
5 %
U
T
(>95% dip in
U
T) for 5
sec
<5 %
U
T
(>95 % dip in
U
T) for
0.5 cycle
40 %
U
T
(60% dip in
U
T) for 5
cycles
70 %
U
T
(30% dip in
U
T) for 25
cycles
5 %
U
T
(>95% dip in
U
T) for 5
sec
Mains power quality
should be that of a typical
commercial, hospital, or
residential environment. If
the user of the
ERO•SCAN
®
requires
continued operation
during power mains
interruptions, it is
recommended that the
ERO•SCAN
®
be powered
from an uninterrupted
power supply.
Power Frequency (50/60
Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
Note:
U
T is the a.c. mains voltage prior to application of the test level.
Summary of Contents for ERO SCAN