Operating Instruction
ERO•SCAN
®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
65
8106556-5 02/15
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The ERO•SCAN
®
is intended for use in the electromagnetic environment specified below. The customer or
the user of the ERO•SCAN
®
should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - Guidance
RF Emissions
CISPR 11
Group 1
The ERO•SCAN
®
uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B Limits
The ERO•SCAN
®
is suitable for use in all
commercial, industrial, business, hospital, and
residential environments.
Harmonic Emissions
IEC 61000-3-2
Class A Category
Voltage Fluctuations /
Flicker Emissions
IEC 61000-3-3
Complies
Recommended Separation Distances between Portable and Mobile RF
Communications Equipment and the
ERO•SCAN
®
The ERO•SCAN
®
is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the ERO•SCAN
®
can help prevent electromagnetic interferences by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the ERO•SCAN
®
as recommended below, according to the maximum output power of the
communications equipment.
Rated Maximum
Output Power of
Transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d
= 1.17
√𝑃
80 MHz to 800 MHz
d
= 1.17
√𝑃
800 MHz to 2.5 GHz
d = 2.23
√𝑃
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitters, where
P
is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply to all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Summary of Contents for ERO SCAN
Page 2: ......