3.1
Status/Error Messages
Display Messages:
Attach Probe
No probe is detected at the start of a test.
BT Device Not
Found
The paired Bluetooth
®
device cannot be detected. The device may be turned
off or too far away.
BT Error #xxx
There is an error condition with the Bluetooth
®
device. Check the status.
BT Not
Configured
The ERO•SCAN
®
instrument is not paired with any Bluetooth
®
device.
Device not
Responding
The printer is not responding to queries from the instrument.
Due For Service
Indicates calibration of instrument is recommended. Message will appear
upon the calibration due date set in the device. Message appears during
device startup once per day.
Fit Error
Cannot Obtain L
For a DP test, the desired level (L1 or L2) cannot be obtained within allowable
limits. User should refit the probe and retry the test.
Fit Error
Too High
For a DP test, the level of the calibration tone is too high. User should refit the
probe and retry the test.
Fit Error
Too Low
For a DP test, the level of the calibration tone is too low. User should refit the
probe and retry the test.
Limit Error
Overflow error during the calculation of the DFTs for a DP test. User should
repeat the test.
Memory Almost
Full
Saved tests are within 5 tests of the maximum limit.
Memory Full!
The maximum saved test limit is reached. The user will need to clear the
memory before any additional tests can be performed.
Power Low!
The battery charge level is too low for operation. The user must charge the
battery before additional tests can be performed.
Printer Error
Indicates a problem with the printer. Check the printer status.
Printer Paper
Out!
Indicates that printer paper has run out.
Time/Date Error
The clock is checked during power on to ensure it has not lost time and been
reset. In the case of clock reset, this message is shown. The user should set
the correct date/time.
Indicator LEDs (lights):
NOISE / Orange
The indicator labeled ‘NOISE’ provides a visual indication (AMBER) that the
noise level measured during the test exceeds a nominal threshold.
Also used to indicate some error conditions and when the outcome of test is
REFER, NOISY, or NO SEAL.
TEST / Yellow
The indicator labeled ‘TEST’ provides a visual indication (YELLOW) that the
selected test is being performed. This indicator will remain on steady state
during the test function.
READY / Green
The indicator labeled ‘READY’ lets the user know that the instrument is not
currently performing a test function and that it is available to perform a test
function.
CHARGE / Blue
The indicator labeled ‘CHARGE’ provides a visual indication (BLUE) of the
battery recharging function and battery status. The rate of illumination of the
indicator provides a means of identifying the status of the charging function.