Cleaning and Maintenance
20.1
Cleaning and Disinfection
Use new eartips for each patient. Eartips are for single patient use only. The probe tube,
which does not make direct contact with the patient, should be replaced if there is any sign
of contamination or if the test will not progress past the Probe Check phase. Disinfection of
the probe tube between patients is not required.
External parts of the instrument/probe can be cleaned to remove visible particulate
contamination. Do not attempt to insert any object into probe.
This instrument is not designated as a ‘sterile’ device. Wiping with a clean cloth or towel
and a mild non alcohol-based disinfecting solution, provides a suitable form of cleaning and
low-level disinfection of the housing and probe exterior. Repeat this weekly, or as often as
conditions warrant, to prevent a build-up of grime from normal handling and use.
We believe low-level disinfection is appropriate for this type of instrument. This may not
conform to the infection control guidelines of the user’s facility. The disinfection materials
and procedures applied in the users’ facility may be more appropriate for their
circumstances than the methods outlined above (see cautions below). The frequency of
cleaning and disinfecting is dependent on the facility’s risk assessment, usage, and test
environment.
Important:
•
Do not immerse the instrument or probe in fluids or attempt
to sterilize the instrument or any of its accessories.
•
Do not allow any fluid to enter the device.
•
Do not use autoclave sterilization.
•
Do not use alcohol-based disinfectants.
•
Take care not to put excessive pressure on the clear display
window or allow any utensil to puncture the display window
or control panel.
NOTE: Long-term exposure to any disinfecting agents has the potential to alter the material
properties of the plastic housing and labeling of the device.
Always follow the safety and disposal guidelines given by the manufacturer of cleaning and
disinfectant chemicals.
20.2
Maintenance
This instrument requires no regular maintenance beyond routine cleaning and annual
calibration. The probe tube requires replacement only when it becomes clogged.