Message
Solutions
Limit
Error
Overflow error during the calculation of the DFTs for a DP test.
User should repeat the test.
Cycle instrument power.
Contact MAICO for service if the problem persists.
Memory almost full
Saved tests are within 5 tests of the maximum limit. Print or
transfer test result to avoid interruption in testing.
Memory Full!
The maximum saved test limit is reached. The user will need to
clear the memory before any additional tests can be performed.
Power Low!
The battery charge level is too low for operation. The user must
charge the battery before additional tests can be performed.
Printer Error
Indicates a problem with the printer. Check the printer status.
Reset the printer or cycle the printer power.
Printer Paper Out!
Replace the paper roll.
Time/Date Error
The clock is checked during power on to ensure it has not lost
time and been reset. In the case of clock reset, this message is
shown. The user should set the correct date/time.