11. Printing is not possible.
12. The data for transmission to the
PC are not current.
13. Repetitive accuracy is outside
of the tolerance.
14. The calibration was all right, but
the probe display an incorrect
probe constant after the
calibration.
15. Measuring error when
conducting a perpendicularity
error measurement.
Perpendicularity error
measurement is not possible
16. No characteristics are show in
the display
17. Workpiece zero point 02 and/or
03 cannot be set!
18. SCALE REF-MARK MISSING
Does not travel to the reference
point!
19. Speed error!
Incorrect settings
Check that the printer has enough
paper in the printer.
Improper contacting (impact)
Probe/workpiece is dirty
Variation in temperature
Probe incorrectly calibrated
Not a standard probe
Probe is not sufficiently clamped in
mount
Battery is almost empty
The distance to gage blocks for
groove and ledge is wrong
Operator correction data is wrong
Dial Indicator or the incremental
probe is not ok.
Display is hidden
Error message „zero point 02 or 03
must be larger than 02 or 03“
If when switching on there is no
probe holder attached, i.e. the
weight balance is incorrect, the
carriage will slide automatically
upwards.
If an obstacle is in the way, then it
cannot travel to the reference point.
The carriage has be moved too
quickly > than 600 mm/sec.
Implement settings in menu 6.13
Data and printer.
Insert more paper, check whether
there is a paper blockage, if so
remove paper. Use a RS232 or
USB data connection cable
Remove the USB connection cable
and reconnect – data will be
updated.
Re-calibrate the probe
Clean the probe/workpiece
Conduct measurements in an
acclimatized room
Switch on temperature
compensation
Check charge status of the battery,
if necessary recharge.
Refer to chapter 6.14.3 „probe
calibration parameters“ adjust the
settings value groove and ledge.
Alter the operator correction data or
set to the works correction
Check the dial indicator or
incremental probe
Press the
key– the values will
reappear.
Workpiece zero point 02 / 03 may
only be set once 01 has been set.
The characteristic for the workpiece
zero point 02 / 03 are before the
characteristic for the workpiece
zero point 01 is in the
characteristics list.
Mount the probe in the holder and
switch the height measuring
instrument off and then back on
again, the reference point will be
found.
Remove obstacle and start again.
Restart the height measuring
instrument with a new reference
point
Problem
Reason
Solution
Summary of Contents for Digimar 817 CLM
Page 8: ...8 Mahr GmbH Digimar 817 CLM 1 2 Unpacking 1 2 N I C K L M O P Q...
Page 10: ...10 Mahr GmbH Digimar 817 CLM 5 350 mm 25 kg 600 mm 30 kg 1000 mm 35 kg...
Page 11: ...11 Mahr GmbH Digimar 817 CLM 6 45 7 8 90...
Page 12: ...12 Mahr GmbH Digimar 817 CLM 9 11 G H 10 12 13 M...
Page 13: ...13 Mahr GmbH Digimar 817 CLM B 3x L M 19 17 18 P D E K N 14 15 16...
Page 14: ...14 Mahr GmbH Digimar 817 CLM 100 240 V 50 60 Hz I 12 h 21 22 Euro US UK 20 23...
Page 152: ......