EN | 29 | ACX120/220 | Maintenance
© MAHLE
8.
Install the new filter paying
attention to the position of
gaskets and to the direction
of the arrow indicating the
fluid flowing direction.
9.
Screw the two connection
nuts of the filter.
10.
Replace the front panel
11.
Carry out the automatic leak
test requested by the
software when switched on
again after the filter
replacement.
14.12
System info
In the main menu select MAINTENANCE
and press “INFO”.
In the INFO page, at any time, the
software version and the serial number
can be displayed.
14.13
System update
The firmware (software) can be
updated by way of a USB stick.(where
the new version is stored)
1.
Insert the USB stick in
USB port (fig.3 pos.4)
2.
Power on ACX120/220.
3.
In the main menu select
MAINTENANCE and
press “SYSTEM
UPDATE”.
4.
A message will appear
that the unit is loading an
update.
5.
The unit may load an
updated language file and
configuration file while
updating.
6.
Once unit is updated, the
software version string on
the introduction screen
during power up will
change
14.14
Maintenance of printer
(optional)
To change the roll of paper follow
instructions below:
1.
Open the lid of the printer as
shown
.
2.
Position the roll of paper inside
the housing in the rotation
direction indicated in the picture;
3.
Pull the paper out of the housing
as indicated in the picture and
close the lid ;
4.
The printer is ready for printing.