20
48-635 E3 Modulevel
®
Displacer Level Transmitter
2.6.5.2
Measurement Type: Interface Level
Display
Action
Comment
1
*Status*
*IfcLvl*
*% Out *
* Loop *
Transmitter Display
2
IfcLevel
xx.xx lu
Transmitter Display
(Alternate Home Menu)
3
% Output
xx.x %
Transmitter Display
(Alternate Home Menu)
4
Loop
xx.xx mA
Transmitter Display
(Alternate Home Menu)
5
LvlUnits
(select)
Select the level units of
measure
Select cm, inches, feet, meters
6
Oper Temp
xxx F
Enter the Process
Operating Temperature
Adjusts factory calibration for actual temperature
(limited by maximum temperature rating of model)
7
Set 4mA
xx.xx lu
Enter the IfcLevel value for
4 mA set point
Assigns Level value to 4 mA loop value
8
Set 20mA
xx.xx lu
Enter the IfcLevel value for
20 mA set point
Assigns Level value to 20 mA loop value
9
Lvl Ofst
xxx.xx lu
Enter desired output when
level is at the calibrated zero
reference
Minimum offset = -(displacer length)
Zero reference is typically bottom of displacer or centerline of
bottom-side process connection.
10
Damping
xx s
Enter time constant of
desired damping
0 to 45 sec
11
Fault
(select)
Select loop current value in
presence of a fault
Select from 3.6 mA, 22 mA or Hold
12
Poll Adr
xx
Enter HART polling address
number
If multidrop is not in use, poll address must = 0
13
Loop Mode
(select)
Enable/Disable
Loop Current Mode
If enabled, loop current follows the PV; If disabled, loop current is
fixed (i.e., multidrop mode).
14
Trim Lvl
xx.xx lu
Enter value to adjust Level
reading
Fine tune IfcLevel reading
15
Trim 4
xxxx
Fine tune 4 mA point
Adjust setting (0 – Trim 20 value) to output exactly 4.0 mA
16
Trim 20
xxxx
Fine tune 20 mA point
Adjust setting (Trim 4 value – 4095) to output exactly 20.0 mA
17
Loop Tst
xx.x mA
Enter a mA output value to
test loop
18
Capture
4.00 mA
Pressing
in Data Entry
Mode captures current level
output as 4.00 mA Set Point
Raise interface level to desired 4 mA point on the displacer.
Press
and
to “capture” this liquid level as your 4 mA
point.
19
Capture
20.00 mA
In Data Entry Mode, can
adjust loop value to corre-
spond to current level output
for setting 20.00 mA Set Point
Set second level point between 8.00 mA and 20.00 mA. Raise
interface level to a second level point on the displacer. Press
and
to “capture” this interface level as your upper loop
value between 8.00 and 20.00 mA.
20
New Pass
xxx
Enter new password (0 – 255) Displays encrypted value of present password
Default value = 0
21
Language
(select)
Select from English, Spanish,
French, German
Language choice for LCD display
22
E3 ModHT
Ver 1.0
Transmitter Display
Product identification
Firmware version
23
DispFact
(select)
Select “Yes” to display factory
parameter menu as below
Selecting “No” takes you back to beginning of menu