2.3  Safety instructions – please read carefully

This symbol means that there is a risk of electric shock.  
An electric shock could cause death or personal injury.

The HT Nucleofector™ system has been certified by international  
safety standards and is safe to use when operated in accordance with 
this manual. 

Only use the HT Nucleofector™ power supply in connection with the  
HT Nucleofector™ plate handler. Under 

no

 circumstances should it be  

connected to any other device delivering high-voltage electrical impulses. 
The system is designed to deliver variable high-voltage impulses for the 
purpose of introducing substrates into eukaryotic cells. 

These electrical impulses can be deadly.

Therefore, use this system with care and take the following  
precautions: 

 

– Only use the system once you have read and understood the  

HT Nucleofector™ manual. The manual should be accessible to all users. 
Make sure that each potential user reads and understands it.

 

– Set up the system in a dry place. Avoid spilling liquid onto or into the 

HT Nucleofector™ power supply or the HT Nucleofector™ plate handler. 
Do 

not

 use the system if any components are wet.

 

– Place the HT Nucleofector™ system components on top of a safe, level 

and stable table or bench.

 

– Before connecting/disconnecting any of the interface cables between 

the HT Nucleofector™ plate handler and the HT Nucleofector™ power 
supply, ensure that the HT Nucleofector™ power supply is switched 
off and the power cable is removed from the power inlet.

 

– Prevent sudden impacts and vibration while moving and transporting 

system components.

 

– Only use HT Nucleofector™ devices in conjunction with the HT 

Nucleofector™ software.

 

– Standard Nucleofector™ solutions and the Nucleofector™ solutions  

used for the HT Nucleofector™ are not compatible.

 

– Do not use solutions or plates from any source other than Lonza.

 

– Unpack the 384-well Nucleocuvette™ plates immediately before the 

experiment. Make sure that the outer contact areas are dry. 

 

– Always place the lid onto the 384-well Nucleocuvette™ plate before 

transferring it to the HT Nucleofector™ plate handler.

 

– Safety may be compromised if any fluid has been spilled in the close 

vicinity of the HT Nucleofector™ system. Ensure that no fluid comes 
into contact with or enters the system components. If any fluid has 
been spilled in the close vicinity of or onto the HT Nucleofector™ plate 
system, contact Lonza scientific support for advice on precautions to 
be taken before further use. 

 

– Do 

not

 open the device housing. Under 

no

 circumstances should circuit 

components be interfered with, as they can deliver an electric shock 
even when the system is not in operation. 

 

– Do 

not

 alter the device in any manner. 

 

– Do 

not 

use the device if the insulation of the high voltage cable 

connecting the plate handler and the power supply is damaged.

 

– Do 

not

 use the device if any of the housing parts or covers are missing.

 

– Do 

not

 detach any of the housing parts or covers.

 

– Do 

not

 expose the device to a humid environment. 

 

– Do 

not

 expose the device to direct sunlight or place the instrument 

in a hot environment.

 

– Under 

no circumstances

 plug the high-voltage cables of the  

HT Nucleofector™ plate handler into external electrode sockets of  
any device other than the HT Nucleofector™ power supply. Use of a 
device from any source other than Lonza will invalidate all warranty  
and liability claims.

 

– Do 

not

 use lids other than those provided with the 384-well 

Nucleofector™ kits as the system is calibrated to the supplied lids. 
Other lids may cause damage to the system. 

 

– Do 

not

 obstruct any moving parts, e.g. the plate carousel of the  

HT Nucleofector™ plate handler.

 

– Do 

not

 move the HT Nucleofector™ plate handler if a 384-well 

Nucleocuvette™ plate is inside the device. Moving the plate handler 
may cause sample fluid from the plate to spill into the device. 

 

– Never place any foreign object onto the device or onto the plate 

carousel. If a foreign object has entered the HT Nucleofector™ plate 
handler, the safety of the device may be compromised. Contact 
Lonza scientific support for advice on precautions to be taken before  
further use. 

 

– The device is not approved for use in fire- or explosion-endangered 

areas, or for use with inflammable or explosive media.

 

– If any component of the HT Nucleofector™ system has been damaged, 

ensure that the HT Nucleofector™ system cannot be used (e.g. by 
disconnecting the HT Nucleofector™ power supply), and contact Lonza 
scientific support for assistance.

 

– Service shall only be performed by personnel authorized by Lonza.

 

– Handling of device parts that may be contaminated by samples shall 

always be performed with protective gloves and any disposal of such 
parts must be according to federal, state or local procedures for clinical 
waste handling and disposal. Use secure leak-proof containers and 
avoid unprotected handling of such parts.

Note: Lonza disclaims all warranties expressed or implied and shall in no event be liable 
for any kind of damages caused by or arising out of any operation or use in violation of the 
safety and handling instructions above. 

2

Summary of Contents for Amaxa HT Nucleofector

Page 1: ...BioResearch Amaxa HT Nucleofector hardware manual For research use only...

Page 2: ...e non transferable right to use the system as well as Lonza s proprietary Nucleofector technology only for research conducted by the buyer whether the buyer is an academic or for profit entity For fur...

Page 3: ...em components 8 2 6 System set up instructions 10 2 7 Instructions for using the HT Nucleofector system 11 3 Running an experiment 12 3 1 Processing a 384 well Nucleocuvette plate 12 3 2 System shutdo...

Page 4: ...provides cell type specificoptimizedprotocolsformanydifferentcelllinesandprimarycells Transfection of any cell With over 160 protocols for cell lines and primary cells optimized by Lonza sR Dteamandm...

Page 5: ...eofector system meets your screening requirements Experience the benefits of the HT Nucleofector system Automation compatible full integration into liquid handling platforms Fast process 384 samples i...

Page 6: ...esearch conducted by the buyer whetherthebuyerisanacademicorfor profitentity Forfurtherdetails about the license refer to chapter 5 For license information contact Lonza Cologne GmbH Phone 49 221 9919...

Page 7: ...eonprecautionsto be taken before further use Donotopenthedevicehousing Undernocircumstancesshouldcircuit components be interfered with as they can deliver an electric shock even when the system is not...

Page 8: ...fector solutions in a biohazard container Refer to your local waste management organization and to the relevant laboratory safety instructions for proper disposal practices 2 5 HT Nucleofector system...

Page 9: ...upply 7 8 10 9 12 11 HT Nucleofector plate handler The HT Nucleofector plate handler serves as the contacting unit of the system Transfection of the samples occurs in 384 well Nucleocuvette plates tha...

Page 10: ...technical specialist Follow the instructions below if the instrument is moved from its original location ConnecttheHVportoftheHTNucleofector powersupply 10 tothe HVportontheplatehandlerUnit 13 usingth...

Page 11: ...m execution optional Software manual 10 Remove 384 well Nucleocuvette plate from the plate handler and transfer samples to appropriate culture plates Optimized protocol 11 Check result file on the com...

Page 12: ...re that plate position A01 is in the upper left corner Place the plate gently into the plate handler retainer position 1 Afterloadingthe384 wellNucleocuvette plate starttheexperimentfrom the computer...

Page 13: ...this may negatively influence the viability of the cells after Nucleofection For detailed recommendations on passage number cell density and confluency refer to the relevant optimized protocol Low tr...

Page 14: ...tch of the power supply 2 Disconnect the main power cable from the power supply power inlet 3 Remove the small plastic cap on top of the plate handler housing see Figure 4 1 to access the emergency li...

Page 15: ...esentations and warranties Lonza Cologne represents and warrants that it is not aware of any legal deficiencies of the patents and that to its knowledge it has the right to dispose of the patents Lonz...

Page 16: ...50 45 cm 5 3 19 6 17 7 in Manufacturing date The year of manufacture is indicated by the second and third digit of the serial number e g a device with the serial number x11xxxxx was manufactured in 20...

Page 17: ...17 Notes 6...

Page 18: ...18 BioResearch HT Nucleofector hardware manual Notes...

Page 19: ...19 Notes...

Page 20: ...ed Use in any other application requires a license from Evrogen To obtain such a license please contact Evrogen at license evrogen com The information contained herein is believed to be correct and co...

Reviews: