Lonza Amaxa HT Nucleofector Hardware Manual Download Page 14

14

BioResearch
HT Nucleofector™ hardware manual

Figure 4.1: Plastic cap covering the emergency release of the contact bridge

Figure 4.2:

 Lift contact bridge by turning the screw clockwise

Figure 4.3: Turn carousel manually to recover samples

4.2   Mechanical  failures

Great care has been taken to ensure that mechanical failures will not occur 
as long as the system is operated using the guidelines presented in this 
manual and in the HT Nucleofector™ software manual. 
In case of a system failure while a 384-well Nucleocuvette™ plate is still 
inside the plate handler unit, proceed as follows to recover your samples:

1.  Turn off the HT Nucleofector™ using the main switch of the  

power supply.

2.  Disconnect the main power cable from the power supply power 

inlet.

3.  Remove the small plastic cap on top of the plate handler housing 

(see Figure 4.1) to access the emergency lift screw of the contact 
bridge.

4.  Using a screwdriver, turn the screw clockwise as far as possible.
5.  The plate handler carousel can be turned manually and samples can 

be recovered.

6.  To recover a 384-well Nucleocuvette™ plate loaded on plate retainer 

position 1 (position 2 is visible at the front of the plate handler) 
turn the carousel counterclockwise to recover your samples. To  
recover samples from plate retainer position 2 (position 1 is  
visible), turn the carousel clockwise.

7.  Remove the 384-well Nucleocuvette™ plate from the plate handler 

carousel.

8.  Contact Lonza’s technical service for analyzing the error and for 

repair of the system.

Summary of Contents for Amaxa HT Nucleofector

Page 1: ...BioResearch Amaxa HT Nucleofector hardware manual For research use only...

Page 2: ...e non transferable right to use the system as well as Lonza s proprietary Nucleofector technology only for research conducted by the buyer whether the buyer is an academic or for profit entity For fur...

Page 3: ...em components 8 2 6 System set up instructions 10 2 7 Instructions for using the HT Nucleofector system 11 3 Running an experiment 12 3 1 Processing a 384 well Nucleocuvette plate 12 3 2 System shutdo...

Page 4: ...provides cell type specificoptimizedprotocolsformanydifferentcelllinesandprimarycells Transfection of any cell With over 160 protocols for cell lines and primary cells optimized by Lonza sR Dteamandm...

Page 5: ...eofector system meets your screening requirements Experience the benefits of the HT Nucleofector system Automation compatible full integration into liquid handling platforms Fast process 384 samples i...

Page 6: ...esearch conducted by the buyer whetherthebuyerisanacademicorfor profitentity Forfurtherdetails about the license refer to chapter 5 For license information contact Lonza Cologne GmbH Phone 49 221 9919...

Page 7: ...eonprecautionsto be taken before further use Donotopenthedevicehousing Undernocircumstancesshouldcircuit components be interfered with as they can deliver an electric shock even when the system is not...

Page 8: ...fector solutions in a biohazard container Refer to your local waste management organization and to the relevant laboratory safety instructions for proper disposal practices 2 5 HT Nucleofector system...

Page 9: ...upply 7 8 10 9 12 11 HT Nucleofector plate handler The HT Nucleofector plate handler serves as the contacting unit of the system Transfection of the samples occurs in 384 well Nucleocuvette plates tha...

Page 10: ...technical specialist Follow the instructions below if the instrument is moved from its original location ConnecttheHVportoftheHTNucleofector powersupply 10 tothe HVportontheplatehandlerUnit 13 usingth...

Page 11: ...m execution optional Software manual 10 Remove 384 well Nucleocuvette plate from the plate handler and transfer samples to appropriate culture plates Optimized protocol 11 Check result file on the com...

Page 12: ...re that plate position A01 is in the upper left corner Place the plate gently into the plate handler retainer position 1 Afterloadingthe384 wellNucleocuvette plate starttheexperimentfrom the computer...

Page 13: ...this may negatively influence the viability of the cells after Nucleofection For detailed recommendations on passage number cell density and confluency refer to the relevant optimized protocol Low tr...

Page 14: ...tch of the power supply 2 Disconnect the main power cable from the power supply power inlet 3 Remove the small plastic cap on top of the plate handler housing see Figure 4 1 to access the emergency li...

Page 15: ...esentations and warranties Lonza Cologne represents and warrants that it is not aware of any legal deficiencies of the patents and that to its knowledge it has the right to dispose of the patents Lonz...

Page 16: ...50 45 cm 5 3 19 6 17 7 in Manufacturing date The year of manufacture is indicated by the second and third digit of the serial number e g a device with the serial number x11xxxxx was manufactured in 20...

Page 17: ...17 Notes 6...

Page 18: ...18 BioResearch HT Nucleofector hardware manual Notes...

Page 19: ...19 Notes...

Page 20: ...ed Use in any other application requires a license from Evrogen To obtain such a license please contact Evrogen at license evrogen com The information contained herein is believed to be correct and co...

Reviews: