D9U003VB3-0101_04
52/65
11.10
Manual Defl ating - Body Zoner (only Virtuoso Pro)
To defl ate individual cell:
Turn selector to the position .
To infl ate individual cell and set APT Mode:
Turn selector to the position
NOTE Green colour in the circle indicates default mode of corresponding cell.
Fig.
Indication of defl ated
individual cell
Fig. Head Zoner
To defl ate individual cell:
Turn selector to the position
To infl ate individual cell and set APT Mode:
Turn selector to the position or
NOTE Green colour in the circle indicates default mode of corresponding cell.
To infl ate to a static pressure in the individual cell:
Turn selector to the position or
NOTE Green colour in the circle indicates default mode of corresponding cell.
11.11 Head Zoner / Heel Zoner
If a Head/Heel Zoner is fi tted (Virtuoso With Head & Heel Zoner (Air + Alternating Air Base) or Virtuoso Pro), it is possi-
ble to confi gure the head and heel section of the mattress for nursing procedures (e.g. prone positioning, intubation). It
permits setting each cell of the head or heel section separately to one of three (Head Zoner) or one of two (Heel Zoner)
modes.
WARNING!
Risk of injury due to incorrect defl ating!
Never open more than 2 valves next to each other!