63/65
D9U003VB3-0101_04
14 Maintenance
WARNING!
Risk of injury when working on the mattress replacement system!
► Ensure that the mattress replacement system is disconnected from the mains power prior to installati
-
on, putting into service, maintenance and deinstallation.
WARNING!
Risk of injury due to defective mattress replacement system!
► Have a defective mattress replacement system repaired immediately.
► If the defect cannot be repaired, do not use the mattress replacement system.
CAUTION!
Material damage due to incorrect maintenance!
► Ensure that maintenance is performed exclusively by manufacturer´s customer service or by authoris
-
ed service personnel certified by the manufacturer.
► If the defect cannot be repaired, do not use the mattress replacement system.
14.1 Regular maintenance
►
Perform regularly visual check of the product (with delivery note if necessary).
►
Ask service department of the manufacturer for addition of the original spare parts if some product parts are
missing.
►
Ask service department of the manufacturer for replacement of any damaged product parts by the original spare
parts.
►
Check inside and outside of mattress and outside of SCU for mechanical damage and signs of severe wear and
tear.
►
Check if mattress and SCU are working properly.
►
Check external air filter in side of SCU for dust and dirt. If dust or dirt is visible, replace filter.
14.2 Spare Parts
The serialt label is located on the SCU and on the mattress. The serial labels contain information for claims and ordering
replacement parts.
Information about spare parts is available from:
■
Manufacturer´s customer service
■
Sales department
14.3
Safety Technical Checks
WARNING!
Risk of injury due to incorrect safety technical checks!
► Ensure that safety technical checks are performed exclusively by manufacturer´s customer service or
by authorised service personnel certified by the manufacturer.
► Ensure that the safety technical checks are recorded in the service and maintenance log.
Safety technical check of the mattress replacement system must be performed at least once every 12 months.
The procedure for performing the safety technical check is stipulated in EN 62353:2014.
NOTE
On request, the manufacturer will provide service documentation (e.g. circuit diagrams, component part lists,
descriptions, calibration instructions etc.) for service personnel for the repair of ME equipment designated by the manu-
facturer as repairable by service personnel.