51/65
D9U003VB3-0101_04
11.8
Transport Mode (Static and Active)
Static Transport Mode allows disconnecting the mattress from the SCU for transporting the bed. Pressure in the static
part of the mattress stabilizes for 8 hours.
NOTE In Static Transport Mode, alternating pressure therapy is not possible.
To activate Static Transport Mode:
Select Max Mode in order to completely infl ate all mattress cells (see Max Mode - chapter 11. 2. 6).
Disconnect air connector socket and rotate its end and body in opposite directions until red symbol ▲ and
black symbol ● are aligned and no air is escaping.
11.8.1 Active Transport Mode
Active Transport Mode allows the patient being transported in bed to continue to receive ALT or CLP therapy. See Accu-
mulator - chapter 12 (only available on Accumulator Back-up ready SCU).
To activate Active Transport Mode:
Turn on the Accumulator power switch.
Disconnect mains plug from wall outlet.
11.9
Power Failure
In case of a power failure, the mattress will remain infl ated for at least 8 hours. However, alternating pressure therapy is
not possible.
If no power supply will be available for more than 8 hours:
Switch to Active Transport Mode (see Transport Mode - chapter 11.8).
After activation of SCU the mode set before turning off starts again if this previous mode was used for 15 minu-
tes or more. This settings is saved in SCU unless backup memory is completely erased. After that default mode
is set back to APT mode with pressure level 3.
WARNING!
Risk of injury due to incorrect positioning of patient!
Prior to more than 30 minutes of Static Transport Mode, consult qualifi ed hospital staff .
WARNING!
Risk of injury due to power failure!
Alternating pressure therapy is not possible during power failures.
In case of power failure, seek medical advice immediately.
Fig. Transport Mode