11.2.5 APT Mode
Once the mattress is fully inflated, a confirming signal sounds and the mattress switches into APT Mode (Alternating
Pressure Therapy). APT Mode with pressure level 3 becomes automatically default mode if SCU is turned off for a long
time. APT Mode is intended to be used as an aid to patient care in prevention and treatment of skin breakdown related to
pressure damage.
To select APT mode:
Press button
13
once or repeatedly until indicator
4
comes on.
When APT mode is switched on:
3-cell mattress units will inflate and deflate in cycles of 7.5 minutes.
Indicator
4
lights white.
Pressure indicator
3
flashes.
Once the pre-set pressure is reached, indicator
3
will remain solid on.
NOTE After 3 cycles of APT Mode with deflated individual cell an alert indicating fail to inflate is triggered.
11.2.6
Max Mode
Max Mode (maximum inflation mode) provides a solid and stable surface for patient care.
Max Mode will run for 30 min at most. If no other mode is selected within last 5 minutes, a confirmation signal will sound,
and the mattress replacement system will switch back to APT or CLP mode automatically. Selected Max Mode interrupts
the return to the previous mode.
To select Max Mode:
Press button
13
once or repeatedly until indicator
5
flashes yellow.
If Max Mode is required for less than 30 minutes:
Set new mode by pressing button
13
.
If Max Mode is required for more than 30 minutes:
Press MODE Button once or repeatedly until Max Mode is set again after 25 minutes of Max Mode.
When Max Mode is selected:
SCU inflates all mattress cells to the same pre-set pressure.
Indicator
5
flashes white until preset pressure is reached.
Once the set pressure is reached, indicator
5
will remain solid on.
During the last 5 min, an alert signal will sound every minute.
When Max Mode is selected for the second time:
During the last 5 min, an audible reminder will sound every minute to remind the personnel that Max Mode
will soon end automatically.
NOTE It is not possible to select Max Mode a third time. Selecting Max Mode again is only possible when APT Mode or
CLP Mode has been running for at least 30 min.
11.2.7
CLP Mode (Constant Low Pressure Mode)
CLP Mode (constant low pressure mode) keeps the mattress pressure at the selected level. The pressure is checked
every 30 seconds, and adjusted if necessary. CLP Mode is intended to be used as an aid to patient care in prevention of
skin breakdown related to pressure damage.
To select CLP Mode:
Press button
13
once or repeatedly until indicator
14
flashes white.
When CLP Mode is switched on:
CLP indicator
14
will light solid and indicator
3
will flash while pressure is adjusted to the set pressure level.
Once the pre-set pressure is reached, indicator
14
and indicator
3
will remain solid on.