57/65
D9U003VB3-0101_04
Fig.
Run the strap to another D-ring and repeat for the Adjustable end of the strap. (The straps, if required can be fi tted
to the cushion from side to side, or from front to back and adjusted as necessary.)
WARNING!
Risk of injury when seating the patient!
Ensure that seat cushion is completely infl ated!
11.12.5 Initial Operation
Ensure that SCU is not covered and air fl ow around SCU is not obstructed in order to avoid overheating.
Connect mains cable of SCU to mains.
To switch on SCU:
Switch on SCU using power switch on side of SCU.
Indicator
6
fl ashes.
Confi rming signal sounds.
Seat cushion is being infl ated.
During the infl ation process:
Indicator
6
fl ashes white.
When the infl ation process is fi nished:
Indicator
6
lights solid.
Cushion is ready for placement of patient.
NOTE Pressure adjustment is not available in this mode.
If indicator 6 fl ashes during infl ation process and audio alert signal sounds:
Check if air connector is connected correctly.
11.12.6 Use
Preparation
Infl ate seat cushion (see Putting into Service).
Seating of Patient
Sit patient on seat cushion.
For an ideal sitting position
Ensure that chair is suitable for patient’s size in order to guarantee correct seating position.