15.4 Electromagnetic Compatibility
Bed is intended for hospitals except for near active HF surgical equipment and the RF shielded
room of a medical system for magnetic resonance imaging, where the intensity of EM distur-
bances is high.
Bed has defined no essential performance.
List of used cables:
1. Mains cable, maximum length 4 m
2. Handset cable, maximum length 3,6 m
Warning!
It is recommended to avoid the use of this device next to or in block with other de-
vice, because it could lead to improper operation. If such use is needed, this device
and the other equipment should be under surveillance to verify proper operation.
Warning!
Use of the accessories, converters and other cables, than specified and provided
by manufacturer of this bed could lead to increase of electromagnetic emission or
lower the electromagnetic immunity of this bed and lead to improper operation.
Warning!
Mobile RF communication device (including end use devices like antenna cables and
external antenna) should not be used closer than 30 cm (12 inches) from any part
of this bed sentida 03, including cables specified by manufacturer. Otherwise this
could lead to deterioration of functionality of this bed.
Warning!
Do not overload the bed (SWL), respect the duty cycle (“INT.” on the label) and con-
sider chapter Service in order to maintain the basic safety with regard to electromag-
netic disturbances for the expected service life.