39
8.3 Side guard extension for 09 side guards
If you want to attach the extension to the 09 side guards:
•
Pull up the side rails of the nursing home bed to the highest position.
•
Release the clamping lever by turning it 90° and twisting out the mushroom head bolt.
•
Place the side rail extension onto the side rail of the nursing home bed as shown in the
picture above.
•
Twist in the mushroom head bolt and screw the clamping lever tight by turning it 90°.
•
Perform a function test of the side rail, including the assembled side rail extension.
If you want to remove the side guard extension:
• Make the above points in reverse order.
Warning!
The side safety clearance from the head and foot ends must each be approx.
6 cm (2,4 inch). Rule: The distances must be uniformly distributed at both ends. For
nursing home beds in special length (more than 2 m (78,7 inch)) a side guard protec-
tion in special length has to be used.
Warning!
For side rail protections with extensions, please use the available side protection
cushioning (accessory), which is an important additional safety and protection fea-
ture.
ca. 6 cm (2,4 inch) safety clearance