8. Accessories
We recommend using compatible mattresses that have a minimum density of 40 kg/m³ (den-
sity 40). All Linet standard mattresses fulfill this requirement. Please refer to the following over-
view for the suitable mattress dimensions for the corresponding patient surfaces.
Designation of the
patient surface
Dimensions
Aero / Easyclean
12-18 cm (4,7-7,1 inch) × 100 cm (39,4 inch) × 200 cm (78,7 inch)
Comfort
12-15 cm (4,7-5,9 inch) × 100 cm (39,4 inch) × 200 cm (78,7 inch)
Caution!
Do not use water mattresses. In this case the drives or other components can fail and
cause accidents by exceeding the working load. Always observe the safe working
load.
Caution!
Replace Linet mattresses only with equivalent mattresses with identical dimensions.
Also make sure that there are movement slots on the underside of the mattress, as
this may affect the adjustment functions. Do not use spring core mattresses.
8.1 Mattresses
The bed is equipped with a mattress. If necessary, this mattress can be replaced by the same or
a similar product.
Danger!
The use of accessories and combinations of products not authorised for the product
can lead to serious hazards. This is why only accessories released by Linet and pro-
ducts released for combination may be used.