Chapter 5 BERT Function

The analyzer has the function to send a test pattern and the function to compare the received data with the test pat-

tern. By a loopback test or a face-to-face test, the quality of the data communication line including the modem etc. 

can be evaluated and the failure point of the data communication line can be isolated.

5.1 Connection and Settings

Select the measurement port according to the interface to be tested, and connect as follows according to the test 

method.

 

 

2.3 Measurement Port

 

 

Loopback test

The test pattern data sent from this unit is looped back and tested at each point on the communication line. You 

can evaluate the round-trip communication line that goes through each loopback point and can isolate the failure 

point by changing the loopback point

  

Setting and switching of the loopback point in the device under test such as a model may be possible using the self-test 

function of the device under test. If you cannot loop back inside the device, loop back at the connector.

 

 

End-to-End Test

The end-to-end test is a test in which a device with the BERT function as this unit has is connected to the other 

party and the same test pattern is sent and received each other. This test allows you to evaluate the transmit line 

and receive line separately.

Summary of Contents for LE-2500XR

Page 1: ...Ver 1 2020 07 INSTRUCTION MANUAL LE 2500XR LE 3500XR MULTI PROTOCOL ANALYZER...

Page 2: ......

Page 3: ...ers USER LIMITATION This product has not been developed for the use that needs exclusively high reliability and safety aerospace ap paratus trunk communication apparatus nuclear control apparatus medi...

Page 4: ...pair the line monitor This may result in an injury electric shock and ignition Turn off the power of analyzer and unplug the cables immediately when emanating smoke odor or sound Continuous use may re...

Page 5: ...tronics TV or radio Near a wireless station for mobile communications or a specified low power radio station Remove the battery from the analyzer when you throw away Caution Please follow the instruct...

Page 6: ...27 2 8 Connect to the Target Devices 29 Connect to RS 232 29 Connect to RS 422 RS 485 30 Connect to TTL 31 Expansion Interfaces 32 Chapter 3 Monitor Function 33 3 1 Online Monitoring ONLINE 33 Settin...

Page 7: ...Configuration Function 95 How to use 95 6 5 Auto Save Function 96 Preparation and settings 96 Start measurement 97 Check files by Wi Fi connection 98 End of measurement 98 6 6 Auto Backup Function 98...

Page 8: ...T and ESC at the same time Press SHIFT ESC 1 2 Unpacking When you unpack the product check the items below The product has not been damaged during the transit You have received all the standard access...

Page 9: ...to communicate through various interfaces Auto save function which records measured data into SD card or USB flash drive for a long time Timing waveform measurement function which is useful to find ti...

Page 10: ...e interface sub board 6 RS 232C Port A measurement port for RS 232C V 24 7 TTL ExternalSignal Oport Measurement port for TTL An I O port for external signal 8 RS 422 485 Port Measurement port for RS 4...

Page 11: ...l VB Voltage Level on the RS 422 RS 485 Terminal VT TTL Level VIL Minimum of the threshold of input level H VIH Maximum of the threshold of input level L 1 Measuring Pause 2 Information of measured da...

Page 12: ...isplay while measuring Scroll one data line Move the cursor on the condition setting Scroll one character of data Change the setting on the condition setting ENTER Apply selected settings 0 F Enter nu...

Page 13: ...r more recommend 2A or more To replace the battery purchase LINEEYE Lithium ion battery model P 26LS1 and do not use any other Charge the Battery Lithium ion battery is charged while analyzer runs by...

Page 14: ...displayed at the opening screen please contact LINEEYE distributors or LINEEYE directly Self check OK should be displayed If Firmware loader is displayed insert the interface board correctly Before u...

Page 15: ...Menu Chapter 7 Save and Load of the Data 2 3 Measurement Port Press MENU then touch Interface Select the measurement port of target device Also it is able to select the item by and change the setting...

Page 16: ...control and timing of data transmission for simulation 4 3 Auto Line Control Driver control for RS 422 485 only Select ON OFF of driver IC control for simulation 4 2 RS 422 485 Driver Control For TTL...

Page 17: ...ction Press Menu then touch Function Select the operation mode and function Touch Advanced settings for addition settings For Simulation function touch Tx data registration to register transmission da...

Page 18: ...ronous Communication which uses the start and stop bit Ex PC COM port UART communication SYNC BSC Character SYNC Communication which uses 1 or 2 bytes of SYNC characters Ex BSC JCA procedure etc HDLC...

Page 19: ...ilo E represents mega and C is for inserting comma M k and are also appeared on the right bottom of screen Exp To set 945 6kbps input 9 4 5 C 6 D TXD speed Set the communication speed of transmission...

Page 20: ...to transmissions data during simulation For received data stop bit is always 1 bit BCC Set a block check code If None is set block check is not executed Selection items of BCC are changed by total bi...

Page 21: ...e end code 1 or 2 characters in HEX Initial setting is blank If setting nothing in here frame end is judged by Frame end time SYNC BSC Character SYNC Data code Same as ASYNC Data bit Set a data bit le...

Page 22: ...ppress code Set a suppress code in HEX When codes set in this section are monitored continuously the second byte and after ones are not recorded in the capture buffer Bit sequence Set a bit sequence F...

Page 23: ...Packet Set a translation specification of packet level The initial value is X 25 9 5 Translation Display Specification Idle mode for simulation Select an idle state Mark or Flag between frames Leadin...

Page 24: ...t second byte is recognized as CRC error check code Data structure of ASCII Begin code 3Ah Data 0 30h 9 39h A 41h F 46h End code CR LF 0Dh 0Ah Parity Set a parity bit For Modbus protocol it automatica...

Page 25: ...changed 2 8 Connect to the Target Devices For SPI protocol select TTL 5 0V TTL 1 8V from Interface Measurement port and select Output type CMOS Burst Data code Select a data code from ASCII EBCDIC EBC...

Page 26: ...g through the communication channel is received is recorded in the capture buffer and displayed Set on off of recording time stamp and its resolution Line state Line state Mark on the box to record th...

Page 27: ...rded in the buffer memory from being inadvertently overwritten A pop up message will be displayed when starting measurement or reading data from files Do no mark on the box when it uses Auto save func...

Page 28: ...6 hours until full charge if analyzer is turned off If analyzer is turned on it becomes Slow charge Fast charge It takes about 3 5 hours until full charge if analyzer is turned off It takes about 6 ho...

Page 29: ...SSID SSID of wireless access point Password Security key encryption key of wireless access point Input characters are changed to for security purpose Touch the tab of Wireless LAN The initial setting...

Page 30: ...xxxx xxxxxxxx is the serial number Security Port number Set the port number of analyzer The initial value is 10101 Touch Apply at the end If the analyzer can connect to the wireless access point succe...

Page 31: ...pin Dsub9pin Dsub9pin Dsub9pin male male female female male 8 1 1 1 3 2 2 2 2 3 3 3 20 4 4 4 7 5 5 5 6 6 6 6 4 7 7 7 5 8 8 8 22 9 9 9 For monitoring Connect to RS 232 To measure RS 232C select RS 232C...

Page 32: ...is connected as one of the RS 485 nodes If the analyzer is connected as left figure set the analyzer to be DTE for simulation If the analyzer is placed on the termination of the line without equipment...

Page 33: ...lyzer O indicates an output from the analyzer Example of connection for I2C Signal Pin No Lead wire Input Output Signal of target device Monitor Simulation Monitor Simulation SDA 1 Brown I I O SDA SDA...

Page 34: ...oard is available at LINEEYE web page Write the firmware into the analyzer using le8firm exe which is stored in the attached CD Once writing the firmware into the analyzer appropriate firmware will au...

Page 35: ...RUN Then data will be recorded in the capture buffer while displaying data on the screen Transmission and reception data is displayed in two lines as one set is the mark which indicates the place of n...

Page 36: ...7bits or more of 1 is continuously detected 1 Shart Frame frame length is short Block check code normal BCC or FCS is normal Block check code abnormal BCC or FCS is abnormal HDLC starting flag patter...

Page 37: ...rd control Additional info The time until reaching to be ASYNC or receiving a specific address in HDLC is considered to be idle state with without data existence If the time of one bit is slower than...

Page 38: ...ter the measurement 6 3 Timing Waveform Measurement Function End of Measurement Press STOP to stop measuring It is possible to stop measuring automatically under the specific conditions by the trigger...

Page 39: ...Top page Display the oldest data positioning number 0 END X Key or touch To Last page Display the latest data It is able to mark the specific data and return to that position any time Touch Set to mak...

Page 40: ...me type in the form of mnemonic NS Displays the frame sequence number with the decimal notation PF Displays the logical value of P F bit NR Displays the frame sequence number with the decimal notation...

Page 41: ...s the protocol identifier with the hexadecimal notation Message Displays the contents separately for message type in the form of mnemonic CRF Displays the value of nominal number flag CR Displays the...

Page 42: ...the specification of Translation display anytime during test too Dump Display one frame per a line in HEX MODBUS Translate the address and control part on top of the frame Item Meaning Time stamp Sho...

Page 43: ...h 61h After 1 bit shift Received data 11001100 01001100 10000011 01000011 11000011 IR 33h Sy 32h A C1h B C2h C C3h Mark bit added Bit Shift Function This function shifts the separation of characters o...

Page 44: ...when it is measured Specifying data with time stamp enables to find data when communication trouble occurs In the right figure it retrieves the data between 22 46 30 on Feb 18th and 22 47 00 on Feb 1...

Page 45: ...of each signal as same as start condition On state indicates the active state of signal For RS 232C it is 3 or higher space state of voltage level For TTL it is a low level For TTL only TXD RXD RTS a...

Page 46: ...TREND LE 3500XR only The statistical analysis function is capable of counting the occurrence of the events such as the number of characters and frames in a specific period and displaying the results o...

Page 47: ...hange the range Scroll Swipe the bar graph or data table to scroll data Resolution Set a resolution unit time for statistical processing on a horizontal scale of a graph 1 240 min Statistically proces...

Page 48: ...on or reception data is selected in a memory through the monitor function And then that data is transmitted as simulation data It is useful to perform a reproducing test for data with the same com mun...

Page 49: ...be registered At the beginning some tables of Group0 are displayed on the screen Touch Group0 on the bottom of screen to change to the group Only the first 27 characters are displayed on the screen S...

Page 50: ...or 17h Press SHIFT ENTER again when you change the Configuration of BCC Data code and data bit Inputted data is displayed with the selected data code Configuration Data code And the data code cannot b...

Page 51: ...rning off the power Delete table Touch Clear table to delete a table Touch Execute or press ENTER To cancel this operation touch Cancel or press ESC Fill in data Touch Fill data and specify the begin...

Page 52: ...ACK0 SHIFT C DC1 11h SHIFT 6 ACK1 SHIFT D DC3 13h ASCII JIS HEX EBCDIC EBCDIK Transcode Others ENQ 05h 2Dh 2Dh ACK 06h 2Eh 3Ch NAK 15h 3Dh 3Dh WACK 10h 3Bh 10h 6Bh 1Fh 26h EOT 04h 37h 1Eh ACK0 10h 30...

Page 53: ...Write request to the slave address 41h 42h 43h 3byte of transmission data A1h 1010000b Re start Read request to the slave address bit 0 1 00h 00h 00h 3byte of dummy data Move the cursor to the A1 the...

Page 54: ...driver will be in disable high impedance state It is used to control the driver IC manually for Manual or Program mode When OP SB10N sub board is inserted to the analyzer and Auto line control is chec...

Page 55: ...the range between 0 and 9999 ms DCE mode Set the timing of following item in the range between 0 and 9999 ms Press MENU and touch Interface and set Mode to DTE or DCE Mark on Auto line control and tou...

Page 56: ...a Press MENU and touch Operation mode Simulation Function MANUAL Touch Advanced settings in the bottom of screen Set the following items below Delay time setting is valid only for ASYNC and ASYNC PPP...

Page 57: ...trol DCD and CTS change their status according to the setting Transmission data is displayed in the RXD line and respond data will be displayed on the TXD line Press SHIFT E F to change control lines...

Page 58: ...9999 1msec unit Select the recording area from BUF1 or BUF2 which is the half of capture memory Monitor communication data and perform a reproducing test in the Buffer mode Monitor data and save data...

Page 59: ...analyzer is DTE mode RTS and DTR will be in an active state at the beginning of test And when the analyzer is DCE mode CTS DSR and DCD will be in an active state Press STOP to stop transmission and si...

Page 60: ...ed control line is non active When the analyzer is DTE mode CTS or DCD set at Watch Operate line is used When the analyzer is DCE mode RTS or DTR set at Watch Operate line is used X ON X OFF code Set...

Page 61: ...ine active immediately If the Initial state is OFF it will make the selected control line active after the period of OFF time When the analyzer is DTE mode RTS or DTR is used When the analyzer is DCE...

Page 62: ...op DTR DSR or DTR DCD Character is valid only for ASYNC protocol set at Configuration setting Touch Advanced settings in the bottom of screen 4 7 Echo back test ECHO mode The echo back test is the fun...

Page 63: ...oard OP SB10N DCE DSUB25pin DTE DSUB25pin SD connected in in ternal circuit RD RD connected in in ternal circuit SD RTS connected in in ternal circuit CTS CTS connected in in ternal circuit RTS DTR co...

Page 64: ...elf station it waits for a next message Item Description Range of Selection ADDRESS Station Address Within eight characters in HEX RESPONSE Delay Time 0 to 99999 1msec unit TABLE No Transmission Table...

Page 65: ...SYNC BSC Parity Error BCC Error HDLC SDLC FCS Error In the master mode analyzer works as a master station And it transmits a polling message to each station address and analyzes returned data Response...

Page 66: ...data In this time from the 1st to 23rd character of registered data will be effectively compared data Error X When communication errors are generated in response message from slave station selected tr...

Page 67: ...Repeat count 0 Not Finished Finished into a stop state Run Next station No Set Adding slave station address to polling message Tr a n s m i t t i n g p o l l i n g message Time out A Y No response Re...

Page 68: ...END TBL command and IF TBL command 4 1 Register Transmission Data Timer There are 2 kinds of timers TM 0 and TM 1 used for IF TM command timeout program and SET TM command start stop command Set the u...

Page 69: ...sor moves to the sub command entry field Desired sub commands will be displayed in the right side of the screen Touch or press 0 7 to select a desired sub command To edit the sub command move the curs...

Page 70: ...It is able to analyze the proto col and sequence of transmission and reception Touch to check the timer counter or register value Amount of traffic and speed of communication changes the delay time b...

Page 71: ...ecified label number if the counter exceeds the setting value IF LN L Branches to a specified label number if the logic values of control lines meet with the setting IF REG REG L Branches to a specifi...

Page 72: ...d stop characters the BCC code is added automatically HDLC SDLC The flag and the FCS code are automatically inserted to data and are transmitted ASYNC PPP If transmission data forms a frame the FCS co...

Page 73: ...mits the data of the transmission data table of the table group specified in the operand field The data to be transmitted corresponds with the key number you press When you specify a data table at whi...

Page 74: ...he frame buffer Ex SEND TBL 00 WAIT FRM CLR Some of the commands such as new SEND SET LN SEND commands cannot transmit data until finishing the previous data transmission In this case the program beco...

Page 75: ...or Position Enter Operation Command Field 2 Sub Command Field 1 Operand Field Select 0 CLR or 1 NOCLR Operation When the WAIT FRM command is executed the dedicated frame buffer is deleted and initiali...

Page 76: ...ed label No GOTO Command Designated Label Number Branch Command The GOTO command branches the program control unconditionally to designated label number GOTO L Entering Cursor Position Enter Operation...

Page 77: ...Position Enter Operation Command Field 4 Sub Command Field 2 Operand Field Enter a timer number with the key 0 or 1 Enter a label number in three digit decimal figures with the key 0 to 9 Operation Th...

Page 78: ...Data Table Comparison Entering Cursor Position Enter Operation Command Field 4 Sub Command Field 6 Operand Field Operand Field Specify the table group in the 1st digit and the table No in the 2nd digi...

Page 79: ...er contents DEC When the program starts all the registers will be set 000 2 SET LN Control Line Logic Setting Command Entering Cursor Position Enter Operation Command Field 7 Sub Command Field 1 Opera...

Page 80: ...ield Enter a counter number with the key 0 or 1 Enter a preset value of the counter in six digits decimal figures with the key 0 to 9 Or specify increment or reset with the key A B Operation The prese...

Page 81: ...ation such as trigger valid invalid and trigger action are not used 6 1 Trigger Function Set more than one INT commands If more than one INT commands are set and they branch to different labels only t...

Page 82: ...the main routine from a subroutine which the program control is branched to by the INT command If the label number of three digits is entered in the operand field the main routine is restarted at the...

Page 83: ...a table 00 WAIT FRM CLR Waits for receiving a signal of one frame SET TM0 STOP Stops timer 0 IF TRG1 L003 Judges errors IF CHR AK L002 Judges AK characters 06h GOTO L001 Branches to retransmission pro...

Page 84: ...n data sent from this unit is looped back and tested at each point on the communication line You can evaluate the round trip communication line that goes through each loopback point and can isolate th...

Page 85: ...es for each time set at Time resolution 1 to 1440 minutes The receive bit is exponential For example 1 0E5 10 000 When testing asynchronous communication Set the protocol ASYNC It is tests at the spee...

Page 86: ...is received and synchronization is established Sync search mes sage is displayed After synchronization is established Measuring message is displayed and error rate measure ment starts When using the R...

Page 87: ...Error bit Number of bit errors 0 9999999 9 99E9 Bit error rate Bit error rate 0 00E 0 9 99E 9 Sync loss Number of loss of synchronization 0 9999 3 Receive block Number of received blocks during syn c...

Page 88: ...olling the timer counter it can measure the communication response time and count the number of occurrences of specific events This is useful when analyzing the flow of communication under the specifi...

Page 89: ...igger condition Target line Select either TXD side or RXD side Character Set the data string to be detected up to 8 characters X Don t Care can be input by END X and of bit mask W0 to 2 and SDLC Setti...

Page 90: ...nter Function Idle time The trigger condition is met when the idle time being measured exceeds the set value Idle time can be set in the range of 0 to 9999 The time unit is as same as the idle time un...

Page 91: ...f saving The amount to be saved at one time can be specified by the number of data maximum 9999 before offset after offset with the trigger point as the center Insert a storage device SD card or USB f...

Page 92: ...communication error occurs the buzzer sounds and the number of error occurrences is counted Enable trigger 0 and 1 and set Error as the factor and Buzzer and Counter control increment as the action Me...

Page 93: ...to 0 at the start of measurement and the status changes to stop 2 It starts stops or restarts start after cleared to 0 according to the control information by the trigger function 3 When the set value...

Page 94: ...lect the sampling clock cycle that is 5 to 10 times faster than the communication speed of the measure ment target from 50nsec 100nsec 200nsec 500nsec 1 sec 2 sec 5 sec 10 sec 20 sec 50 sec 100 sec 20...

Page 95: ...with which it automatically stops the waveform monitor measurement The signals that can be set vary depending on the interface and the option board being used 0 0 1 1 END X X Don t care SHIFT 0 Fallin...

Page 96: ...isplayed at C M on the first line of the screen Search of signal status When you touch Search at the bottom of the screen the status display 1 0 at the cursor position on the left of the screen become...

Page 97: ...starts the measurement If you touch Cancel or press STOP or ESC before it starts measurement the analysis processing of communica tion conditions will be stopped and the current communication conditio...

Page 98: ...file management function and the attached PC link software it is useful to analyze the communication failure of un known cause Preparation and settings Set the SD card or USB flash drive of the appro...

Page 99: ...s due to media limitations Start measurement When starting measurement by RUN if there is an auto save file in the storage device an overwrite confirmation message will be displayed If you touch RUN o...

Page 100: ...ible never remove the stor age device during auto save including when Under cleanup message is displayed 6 6 Auto Backup Function As the capture memory is DDR3 SDR AM the data in the memory disappears...

Page 101: ...ing one or both of Run time or Stop time External power on Run When checked measurement starts 10 seconds after the start of USB bus power supply It is used when you want to start measurement in conju...

Page 102: ...t file of the screen display saved by the SHIFT ESC operation is saved in the storage device specified by Screen shot store at the Display tab of System Settings in MENU 7 2 File Management Function Y...

Page 103: ...ctory screen The measure ment data file extension DT is automatically added The file name can be specified with up to 8 characters You can not enter lowercase letters Save setup When you touch Save se...

Page 104: ...e type and the updating date time file time stamp range The following file types can be specified for the filter display All files Analyzer measurement data file and setting data file Analyzer setting...

Page 105: ...window or press ENTER again to load the file data Files with extensions other than DT and SU cannot be loaded Loading the file in subdirectory On the directory screen touch the subdirectory hereafter...

Page 106: ...trigger function TGSAVEnn DT you can do so by using the filter display of the files Long tap again or press ESC to return to the original display and cancel the selection Error message If an error me...

Page 107: ...2014 53 EU In the other countries the Wi Fi function is not available because of radio law regula tion Installation of the PC link software PC link software has a product version and a light version...

Page 108: ...mea surement data in real time on the screen while recording the mea surement data Click to stop the measurement Remote control By clicking the button on the toolbar the image of the analyzer is displ...

Page 109: ...here is no data is converted to HEX line Time information and line state of control line are converted as follows SX A B EX 02 41 42 03 SX A B EX Display Converted HEX line Character line If character...

Page 110: ...8 01 INFO 1 0 0 G 000654484520515549434B2042524F574E204 SD 2315225714 30 INFO 0 1 1 G 303030330D4142434445460D0A RD 2315225737 01 RNR 0 0 G 0A020004084F4B5F4C452D333530305852 SD 2315225738 01 INFO 1 0...

Page 111: ...2020 03 25 15 42 03 Model LE 3500XR Version 1 00 Extension Standard Serial No 99999999 Start time 2020 03 23 15 41 58 Stop time 2020 03 23 15 42 08 MONITOR DATA X 25 PACKET PROTOCOL HDLC S SPEED 1M R...

Page 112: ...SD 155943 3 Diagnostics G 01 SD 155944 3 Read holding registers G 00B00002 SD 155944 3 Read holding registers G 0400000000 SD 155944 2 Read holding registers G 00B00002 SD 155946 2 Read holding regis...

Page 113: ...by the explorer 3 Select Oversized Icon from Display on the explorer toolbar to check the saved screen image 4 Drag the screenshot file you want to use and paste it in a word or Excel to use as measur...

Page 114: ...analyzer from a PC The library can be downloaded from the LINE EYE website For Windows Windows 7 8 10 VC 6 0 and VC NET are supported For Linux Some distributions are supported Supported partly a fee...

Page 115: ...eives either the ITB code or the character set at End code of BCC in the configuration it finishes the calculation including that character BCC judgment When the calculation end character is received...

Page 116: ...sends data in sync with DCE s RT1 clock Send by external clock Receive by internal clock Send by external clock Receive by internal clock Use of AR Auto Regulation In synchronous communication using...

Page 117: ...1 frame ASYNC Data string until the idle time longer than one set at Frame end time in the configuration is detected or the data set at Frame end code is received SYNC BSC Data string from the synchro...

Page 118: ...I appears EBCDIC 0 1 2 3 4 5 6 7 8 9 A B C D E F 0 NU DL DS 0 1 SH D1 SS a j A J 1 2 SX D2 FS SY b k s B K S 2 3 EX D3 WS IR c l t C L T 3 4 PF RE BP PN d m u D M U 4 5 HT NL LF TN e n v E N V 5 6 LC...

Page 119: ...A LF SB J Z j z B VT EC K k C FF FS L l D CR GS M m E SO RS N n F SI US O _ o DT 0 1 2 3 4 5 6 7 8 9 A B C D E F 0 NU DL 0 P p 1 SH D1 1 A Q a q 2 SX D2 2 B R b r 3 EX D3 3 C S c s 4 ET D4 4 D T d t 5...

Page 120: ...NS F O W 6 7 DT PC EC ET G P X 7 8 GE CN SA S2 H Q Y 8 9 S1 EM SE IT I R Z 9 A RT US SM RF B VT C1 CP C3 C FF IF MA D4 D CR IG EQ NK _ E SO RS AK F SI IB BL SB 0 1 2 3 0 2 1 3 1 k j l 2 s t 3 b a c 4...

Page 121: ...C L T 3 4 D M U 4 5 E N V 5 6 F O W 6 7 G P X 7 8 H Q Y 8 9 I R Z 9 A SX EC SY B C D BL US EQ NK E SB ET EX EM F EB DL HT DT 0 1 2 3 0 1 1 J A 2 2 S K B 3 3 T L C 4 4 U M D 5 5 V N E 6 6 W O F 7 7 X P...

Page 122: ...ieve Not Ready N R P F 0 1 0 1 REJ REJ REJect N R P F 1 0 0 1 SNRM Set Normal Responce Mode 1 0 0 P 0 0 1 1 SNRME Set Normal Response ModeExtended 1 1 0 P 1 1 1 1 DISC RD DISConnect Request Disconnect...

Page 123: ...NR Recieve Not Ready N R P F 0 1 0 1 REJ REJ REJect N R P F 1 0 0 1 SARM DM Set Asynchronous Responce Mode Disconnect Mode 0 0 0 P F 1 1 1 1 SABM Set Asynchronous Balanced Mode 0 0 1 P 1 1 1 1 SABME S...

Page 124: ...SD side RD side SD side RD side b8 b7 b6 b5 b4 b3 b2 b1 INFO INFOmation N S 0 N R P RR RR Receive Ready 0 0 0 0 0 0 0 1 N R P F RNR RNR Recieve Not Ready 0 0 0 0 0 1 0 1 N R P F REJ REJ REJect 0 0 0 0...

Page 125: ...rve Not Ready P R 0 0 1 0 1 REJ REJect P R 0 1 0 0 1 CR IC Call Request Incoming Call 0 0 0 0 1 0 1 1 CA CC Call Accept Call Connected 0 0 0 0 1 1 1 1 CQ CI Clear reQuest Clear Indication 0 0 0 1 0 0...

Page 126: ...ect 0 0 1 0 0 0 0 1 USER INFO USER INFOrmation 0 0 1 0 0 0 0 0 DISC DISConnect 0 1 0 0 0 1 0 1 REL RELease 0 1 0 0 1 1 0 1 REL COMP RELease COMPlete 0 1 0 1 1 0 1 0 REST RESTart 0 1 0 0 0 1 1 0 REST A...

Page 127: ...CPIP Van jacobson Compressed TCP IP 002f VJUTCPIP Van jacobson Uncompressed TCP IP 0031 BPDU Bridging PDU 0033 ST Stream Protocol TS II 0035 VINES Banyan Vines 0039 AT EDDP AppleTalk EDDP 003b AT SB A...

Page 128: ...l 8043 AscomCP Ascom Timeplex 8045 LBLBCP Fujitsu LBLB Control Protocol 8047 DCA CP DCA Remote Lan Network Control Protocol 8049 SDCP Serial Data Control Protocol PPP SDCP 804b SNA802CP SNA over 802 2...

Page 129: ...bits of ACCM as 0 Example When all the ACCMs between communication devices are bit ON 1 As shown in the figure above when the data actually flowing on the line is 7E FF 03 00 21 31 32 11 33 34 7E 11...

Page 130: ...HDLC SDLC X 25 MODBUS Not Supported I2C SPI Burst 2 Expansion Protocol CAN CAN FD Device Net 3 LIN CXPI CC Link 3 1Mbps or less only Capture Memory Memory capacity 4 DDR3 SDRAM 100M Byte 2 split avail...

Page 131: ...ction Logic analyzer function 1KHz to 20MHz 14 steps 4 096 samplings Bit error rate test Measure error rate conforming to ITU T Notification G 821 pa rameter Simulation function Operation Spec Enables...

Page 132: ...85 measurement test port Signal definition of RS 422 485 port 1 I is an input to the analyzer O is an output from the analyzer Name RS 232C V 24 Input Output 3 Line state LED DSUB25 Pin 2 MONITOR DTE...

Page 133: ...er 2 For SLAVE simulation connect the SDO of this unit to the MISO to be tested and the SDI of this unit to the MOSI to be tested 3 When the measurement port is set to TTL and the simulation is execut...

Page 134: ...te LED Dsub25 Pin Dsub15 Pin Dsub37 Pin M type 34 Pin Shield ground FG 1 FG 1 FG 1 FG A Transmission data TXD A 2 T A 2 SD A 4 TXD A P SD TXD B 14 T B 9 SD B 22 TXD B S Receiving data RXD A 3 R A 4 RD...

Page 135: ...SCL Blue ST1 ST2 Reception synchronous clock RXC Purple RT External trigger input TRG IN Gray Signal ground GND Black External trigger output TRG OT White Signal Terminal Input out put Line state LED...

Page 136: ...Clock setting MENU and F Program edit display of the PROGRAM mode If the transition destination setting screen is not valid for the current function or operation mode such as pressing the MENU or 9 ke...

Page 137: ...ck Next of the firmware transfer software click Select to select the firmware file extension FW2 to be transferred and click Next 7 Click Start here to start the transfer When Transmission and writtin...

Page 138: ...he soft reset press ENTER TOP DEL when turning on the power It returns to factory default setting Please note that all data will be erased Line state LED does not light Connect the cable correctly Che...

Page 139: ...l stress misuses abuse negligence or accidents LINEEYE disclaims all other warranties including the warranties of merchantability fitness for some par ticular purpose and noninfringement of third part...

Page 140: ...5XRE LE 4F Marufuku Bldg 39 1 Karahashi Nishihiragaki cho Minami ku Kyoto 601 8468 Japan Tel 075 693 0161 Fax 075 693 0163 URL https www lineeye com Email info lineeye co jp Printed in Japan LINEEYE C...

Reviews: