TruScan 532 Single - Wavelength – Operators Manual Rev. No 02
Page 23 of 114
Safety Filter
Fixed filter OD 4 @ 532nm
Beam
Divergence
Cone angle of 20°
Pattern Types Single, Line, Square (2x2, 3x3, 4x4, 5x5), Macular Arc (double, triple),
Circle (full, 1/4, 1/2, 3/4) and Triangle – Spot width, density and 360
rotation available for each pattern
Laser Indirect Ophthalmoscope (LIO)
Indirect model Keeler All Pupil II
Retinal Spot
size
300µm nominal at focus with 20D Laser Lens
Illumination
power
From Laser Console. Internal PSU allows intensity adjustment.
Connector is on Console Front Panel
Fiber length
3 m
Weight
Less than 500gms without laser attachment
Safety Filter
Fixed filter OD 4
Beam
Divergence
Cone angle 20c
Endo-ocular Probes
Probe Types
Straight, Curved and Aspirating
Fiber Length
3 m
Safety Filters
OD 4 in housing for installation to Operating Microscope before
Laser can be fired
Sterile
Sterilized by Ethylene Oxide and packed as Single use devices ready
for use
Beam
Divergence
0.2 NA
Safety filter
Safety Filters OD 4 @ 532nm