TruScan 532 Single - Wavelength – Operators Manual Rev. No 02
Page 109 of 114
Err 501 - Shutter
Failure
-Shutter failed to open or
close
-Reboot the system
-If the problem persisted, call the
authorized service or manufacturer
for further assistance
Err 502 - Laser
Diode Current Too
High
-Laser diode current is
out of range or over the
limits
-Call the authorized service or
manufacturer for further assistance
Err 503 - Aiming
Laser Failure
-Aiming laser beam failure
or the laser intensity is >
1mW
-Call the authorized service or
manufacturer for further assistance
Err 5041 - Laser
Diode Temperature
Out of Tol/-
2.5C
-Ambient temp. is too high
or low which causes the
laser laser temp. TEC unable
to reach the stable or
preset value within 5 mins.
-Reboot the system
-If the problem persisted, call the
authorized service or manufacturer
for further assistance
Err 5042 - Laser
Sensor Failure
-Laser sensor failure
(opened or shorted
sensor)
-Power off ; call the authorized
service or manufacturer for further
assistance
Err 5043 - Cavity
Temperature Out of
Tol/- 2.5C
-Cavity Temperature
sensor can not reach to
the /- 2.5C
-Reboot the system
-If the problem persisted, call the
authorized service or manufacturer
for further assistance
Err 5044 - Cavity
Temperature
Sensor Failure
-Cavity temp. sensor
failure or the sensor is
open or shorted circuit
-Call the authorized service or
manufacturer for further assistance
Err 5051 - Remote
Communication
Error
-No communication
between remote and the
laser console within the
time of frame of 300ms
-Reboot the system
-If the problem persisted, call the
authorized service or manufacturer
for further assistance
Err 5052 - Remote
Time Up Error
-Timer in remote stopped
running for 1ms
-Reboot the system
-If the problem persisted, call the
authorized service or manufacturer
for further assistance
Err 5053 - Remote
ROM Error
-mismatched code in the
remote software when
-Reboot the system
-If the problem persisted, call the