Program with the MODBUS Protocol
248083440-1 Rev 5
7-5
Location (Register 40026)
For Particle Counters, this value specifies the location where a sample
was recorded.
Hold Time (Registers 40031, 40032)
The Hold Time is used for pausing in between samples for multiple
cycles.
This time is specified in seconds. The maximum value is 359,999
seconds (high word: 5, low word: 32319) which is 99 hours, 59
minutes, and 59 seconds. To set the Hold Time to a value less than 9
hours, 6 minutes, 7 seconds, enter the number of seconds in the
low
register
(40032).
During Hold Time, the Device Status bit is 0 (Idle).
Sample Time (Registers 40033, 40034)
The Sample Time specifies the time period of each sample. This time
is specified in seconds. The maximum value of the sample time is
86,399 seconds (high word: 1, low word: 20863) which is 23 hours, 59
minutes, 59 seconds.
To set the Sample Time to a value less than 9 hours, 6 minutes, 7
seconds, enter the number of seconds in the
low register
(40034).
During the Sample Time, the Device Status is 1 (Sampling).
Summary of Contents for ApexP3
Page 1: ...ApexPortable P A R T I C L E C O U N T E R OPERATING MANUAL ApexP3 ApexP5...
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions ApexPortable Airborne Particle Counter Operating Manual...
Page 6: ......
Page 12: ...Lighthouse ApexPortable Operating Manual ii 248083440 1 Rev 5...
Page 16: ...Lighthouse ApexPortable Operating Manual 1 4 248083440 1 Rev 5...
Page 26: ...Lighthouse ApexPortable Operating Manual 3 6 248083440 1 Rev 5...
Page 30: ...Lighthouse ApexPortable Operating Manual 4 4 248083440 1 Rev 5...
Page 87: ...Operate the ApexPortable 248083440 1 Rev 5 5 57 Figure 5 95 Failed Report Sample...
Page 92: ...Operate the ApexPortable 248083440 1 Rev 5 5 62...
Page 120: ...Lighthouse ApexPortable Operating Manual A 18 248083440 1 Rev 5...
Page 122: ...Lighthouse ApexPortable Operating Manual B 2 248083440 1 Rev 5...
Page 128: ...Lighthouse ApexPortable Operating Manual I 6 248083440 1 Rev 5...
Page 129: ...00...