Operate the ApexPortable
248083440-1 Rev 5
5-5
•
Service Indicator:
Indicates that the instrument may be in need of
service. If wrench displays, please contact a Lighthouse authorized
Service Provider at 1-800-945-5905 (USA Toll Free) or 1-541-770-
5905 (Outside of USA) for assistance or send an email to
•
μ
:
The mu (m-yoo) symbol indicates the particle sizes, in
micrometers, configured for the instrument. These sizes are pre-
configured at the factory. To order an instrument with non-standard
sizes, please contact a Sales at Lighthouse 510-438-0500 (Sales).
•
DATA DISPLAY:
This indicates whether the counts are being
displayed in Differential (Diff) mode or in Cumulative (Cuml)
mode and, if the counts are normalized to ft
3
or m
3
.
•
MODE:
Displays the current mode selected; possible modes are
AUTO, MANUAL, CONC (Concentration) and BEEP. The
exception to this is when a recipe is being run in which case the
MODE area will display number of cycles run versus the total to be
run (x / 6).
•
CYCLES:
Indicates the number of times the count will be taken at
a given location in Auto mode. "1/ 3" indicates the count is the first
of three samples to be recorded at this location. The maximum
number of cycles is 999. When set to 0, the unit will run in Auto
mode continuously until the STOP button is pressed.
•
SAMPLE:
The Sample Time (hh:mm:ss) is the duration of one
counting cycle. The Sample Time will count down on the MAIN
screen when the instrument is in AUTO or Manual mode so the user
can see how much time is remaining in the sample period. In
Concentration mode, the Sample Time will count up to 6 seconds
per cycle.
Note:
If Hold time is
greater than 1 minute, the
pump will stop during that
time. At the end of the hold
time, the pump will restart.
•
HOLD:
Displays the hold time in between cycles. The maximum
hold time is 23 hours, 59 minutes, 59 seconds.
•
RECS:
This displays the current number of records stored in the
instrument and the total number of records that can be stored. The
data buffer is a circular buffer. The instrument can store up to 3000
records. An asterisk (*) will appear in front of counts when the
buffer wraps.
Differential
Cumulative
Normalized ft
3
Normalized m
3
Summary of Contents for ApexP3
Page 1: ...ApexPortable P A R T I C L E C O U N T E R OPERATING MANUAL ApexP3 ApexP5...
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions ApexPortable Airborne Particle Counter Operating Manual...
Page 6: ......
Page 12: ...Lighthouse ApexPortable Operating Manual ii 248083440 1 Rev 5...
Page 16: ...Lighthouse ApexPortable Operating Manual 1 4 248083440 1 Rev 5...
Page 26: ...Lighthouse ApexPortable Operating Manual 3 6 248083440 1 Rev 5...
Page 30: ...Lighthouse ApexPortable Operating Manual 4 4 248083440 1 Rev 5...
Page 87: ...Operate the ApexPortable 248083440 1 Rev 5 5 57 Figure 5 95 Failed Report Sample...
Page 92: ...Operate the ApexPortable 248083440 1 Rev 5 5 62...
Page 120: ...Lighthouse ApexPortable Operating Manual A 18 248083440 1 Rev 5...
Page 122: ...Lighthouse ApexPortable Operating Manual B 2 248083440 1 Rev 5...
Page 128: ...Lighthouse ApexPortable Operating Manual I 6 248083440 1 Rev 5...
Page 129: ...00...