Lighthouse ApexPortable Operating Manual
5-50
248083440-1 Rev 5
Note:
The Minimum
Volumes are measured in the
standard measurement
required by each specific
report.
The number of digits allowed
for area to calculate volume is
limited to a 5 digit value for ft2
and a 4 digit value for m2.
Based on the minimum requirements displayed in Table 5-1, configure
the instrument as follows:
•
Press MAIN to return to the MAIN screen.
•
Press CONFIG to enter the Configuration screen.
•
Press SAMPLE.
Note:
The Sample time
automatically changes to
31:42 (31 minutes, 42
seconds).
•
Press the CYCLES button.
•
Set the CYCLES value for 1 and press ENTER. Note that the
instrument beeps twice and clears the value. This is intentional as
the minimum cycles is 2 and changing to 1 will cause errors.
Entering more than 2 will work, though, if needed.
•
Once the primary settings of Class and Area are entered, the values
are set for sample time/volume and cycles. Unless required by
SOP, there is no need to change these other values.
•
Make sure to add the needed locations and press BACK.
Note:
This is all that is
needed to meet the minimum
requirements to classify the
report.
Increase any of the
configuration parameters in
order to meet additional
reporting needs.
•
Press the MAIN button.
•
Press START and allow the instrument to record the required
sample(s) at the first location.
Figure 5-81 Prompt to Change LOCation
•
When it has finished, it will prompt to change to next location; a
YES response will change the Location but the user must press the
START button when set up and ready in the next Location.
•
Move the instrument to the second location and press START.
Table 5-1
Minimum Volume per Channel Size
Ch Size (
μ
)
Min Vol (ft
3
)
0.3
0.100
0.5
0.200
1.0
0.919
5.0
31.698
Summary of Contents for ApexP3
Page 1: ...ApexPortable P A R T I C L E C O U N T E R OPERATING MANUAL ApexP3 ApexP5...
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions ApexPortable Airborne Particle Counter Operating Manual...
Page 6: ......
Page 12: ...Lighthouse ApexPortable Operating Manual ii 248083440 1 Rev 5...
Page 16: ...Lighthouse ApexPortable Operating Manual 1 4 248083440 1 Rev 5...
Page 26: ...Lighthouse ApexPortable Operating Manual 3 6 248083440 1 Rev 5...
Page 30: ...Lighthouse ApexPortable Operating Manual 4 4 248083440 1 Rev 5...
Page 87: ...Operate the ApexPortable 248083440 1 Rev 5 5 57 Figure 5 95 Failed Report Sample...
Page 92: ...Operate the ApexPortable 248083440 1 Rev 5 5 62...
Page 120: ...Lighthouse ApexPortable Operating Manual A 18 248083440 1 Rev 5...
Page 122: ...Lighthouse ApexPortable Operating Manual B 2 248083440 1 Rev 5...
Page 128: ...Lighthouse ApexPortable Operating Manual I 6 248083440 1 Rev 5...
Page 129: ...00...