Sample Printouts of Standard Reports
Figure 5-96, Figure 5-97 and Figure 5-98 are examples of a Federal
Standard Ft3, ISO 14644-1 and EU GMP 2009 reports.
Note:
If a recipe is
loaded, the name of the
report session will be
printed on the report.
Figure 5-96 Sample Federal Standard ft
3
Report
*************************
Fed Std Ft3
ApexP3
Serial #: 1412105005
Report Name: FS_26_2_D
Targeted Class: 1000
Room Area: 10.76 ft^2
Air Flow: Unidirectional
Room Status: In Operation
Minimum Samples: 5
Minimum Locations: 2
Data Start:
2015/01/26 09:55:57
Data End:
2015/01/26 10:06:59
Sample Volume: 0.100 ft^3
-----------------
Particle Size: 0.3
Loc Samples Avg Conc
LOC001 5 8003.5
LOC002 5 6524.5
Mean: 7264.0
Min: 6524.5
Max: 8003.5
Std Deviation: 1045.8
Std Error: 739.5
95% UCL: 11930.2
Min Vol: 1.467 ft^3
Concen Limit: 3007
Min Volume: 0.100
Classification: FAIL
-----------------
Particle Size: 0.5
Summary of Contents for ApexP3
Page 1: ...ApexPortable P A R T I C L E C O U N T E R OPERATING MANUAL ApexP3 ApexP5...
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions ApexPortable Airborne Particle Counter Operating Manual...
Page 6: ......
Page 12: ...Lighthouse ApexPortable Operating Manual ii 248083440 1 Rev 5...
Page 16: ...Lighthouse ApexPortable Operating Manual 1 4 248083440 1 Rev 5...
Page 26: ...Lighthouse ApexPortable Operating Manual 3 6 248083440 1 Rev 5...
Page 30: ...Lighthouse ApexPortable Operating Manual 4 4 248083440 1 Rev 5...
Page 87: ...Operate the ApexPortable 248083440 1 Rev 5 5 57 Figure 5 95 Failed Report Sample...
Page 92: ...Operate the ApexPortable 248083440 1 Rev 5 5 62...
Page 120: ...Lighthouse ApexPortable Operating Manual A 18 248083440 1 Rev 5...
Page 122: ...Lighthouse ApexPortable Operating Manual B 2 248083440 1 Rev 5...
Page 128: ...Lighthouse ApexPortable Operating Manual I 6 248083440 1 Rev 5...
Page 129: ...00...