Lighthouse ApexPortable Operating Manual
5-46
248083440-1 Rev 5
Note:
If the entered Area
values needs more
locations than available, a
“Not enough locations”
message will be displayed
and the user will be asked
to enter a lesser area.
•
Press the BACK button then the
AREA
button to display the Area
Set Up screen shown in Figure 5-71.
Figure 5-71 Reports Area Set Up Screen
Note:
The minimum area
that can be entered is 6 ft
2
or
1 m
2
.
The maximum area that can
be entered depends on the
number of locations available
on the unit.
With all 200 locations
available, the Max area that
can be entered is limited to a
5 digit value for ft2 and a 4
digit value for m2.
•
Touch the appropriate unit of measure (ft
2
or m
2
) and type in the
area using the keypad. Press ENTER to display the value on the
screen.
•
Press the BACK button.
•
Press the STATUS button to display the Room Status screen shown
in Figure 5-72.
Figure 5-72 Reports Room Status Screen
•
Press the appropriate room status (As Built, At Rest, In Operation)
followed by the BACK button.
Summary of Contents for ApexP3
Page 1: ...ApexPortable P A R T I C L E C O U N T E R OPERATING MANUAL ApexP3 ApexP5...
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions ApexPortable Airborne Particle Counter Operating Manual...
Page 6: ......
Page 12: ...Lighthouse ApexPortable Operating Manual ii 248083440 1 Rev 5...
Page 16: ...Lighthouse ApexPortable Operating Manual 1 4 248083440 1 Rev 5...
Page 26: ...Lighthouse ApexPortable Operating Manual 3 6 248083440 1 Rev 5...
Page 30: ...Lighthouse ApexPortable Operating Manual 4 4 248083440 1 Rev 5...
Page 87: ...Operate the ApexPortable 248083440 1 Rev 5 5 57 Figure 5 95 Failed Report Sample...
Page 92: ...Operate the ApexPortable 248083440 1 Rev 5 5 62...
Page 120: ...Lighthouse ApexPortable Operating Manual A 18 248083440 1 Rev 5...
Page 122: ...Lighthouse ApexPortable Operating Manual B 2 248083440 1 Rev 5...
Page 128: ...Lighthouse ApexPortable Operating Manual I 6 248083440 1 Rev 5...
Page 129: ...00...