Operate the ApexPortable
248083440-1 Rev 5
5-43
REPORT FEATURE
Enable the Report Feature by pressing the
(Default setting)
located at the top left of the screen. The check mark indicates the
feature is enabled, which enables the PRINT MODE status on MAIN
screen. See Figure 5-65 and 5-66.
Note:
If the user sets the
Apex for a minimum volume,
and takes the exact number of
minimum locations and
samples, the user could get a
“Cannot Classify: too few
locations” error if the minimum
volume rounds the sample
time down so that not enough
samples at the minimum
volume are taken.
In this case, add one or two
seconds to the sample time
so that the volume is over the
exact minimum volume.
Figure 5-65 Typical Reports Screen
•
When the REPORT mode is enabled, the currently selected report
type and minimum required samples will be displayed on the
MAIN screen as shown in Figure 5-66.
Figure 5-66 Report Mode Status
•
Depending on the report’s requirements, the number of cycles and
minimum sample time will be calculated when the REPORT mode
is enabled.
•
The number of required locations will be automatically assigned to
the report. The user may change the locations via the report’s
SETUP screen.
•
In this example the display format will be changed to Normalized
and area measurement to ft
3
for FED-209 and m
3
for ISO and EU
GMP. The user may change the display mode in the Reports
Settings screen.
•
The instrument keeps track of and will display on the MAIN screen
the number of samples taken for the report.
Report
Feature
Enabled
Summary of Contents for ApexP3
Page 1: ...ApexPortable P A R T I C L E C O U N T E R OPERATING MANUAL ApexP3 ApexP5...
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions ApexPortable Airborne Particle Counter Operating Manual...
Page 6: ......
Page 12: ...Lighthouse ApexPortable Operating Manual ii 248083440 1 Rev 5...
Page 16: ...Lighthouse ApexPortable Operating Manual 1 4 248083440 1 Rev 5...
Page 26: ...Lighthouse ApexPortable Operating Manual 3 6 248083440 1 Rev 5...
Page 30: ...Lighthouse ApexPortable Operating Manual 4 4 248083440 1 Rev 5...
Page 87: ...Operate the ApexPortable 248083440 1 Rev 5 5 57 Figure 5 95 Failed Report Sample...
Page 92: ...Operate the ApexPortable 248083440 1 Rev 5 5 62...
Page 120: ...Lighthouse ApexPortable Operating Manual A 18 248083440 1 Rev 5...
Page 122: ...Lighthouse ApexPortable Operating Manual B 2 248083440 1 Rev 5...
Page 128: ...Lighthouse ApexPortable Operating Manual I 6 248083440 1 Rev 5...
Page 129: ...00...