Lighthouse ApexPortable Operating Manual
5-52
248083440-1 Rev 5
Data and settings cannot be changed via the instrument web page. The
Apex
Portable
must be set up using the Ethernet interface and correct
IP for the network in which it is installed. Serial-configured (RS485
LAN)
Apex
Portable
instruments do not support the web page feature.
The web page feature requires a valid TCP/IP address to be assigned to
the
Apex
Portable
.
Generate a Report From Data
Figure 5-84 Reports Main Screen
From the MAIN screen, press REPORTS.
1.
Select the Report desired: FED STD FT3, ISO 14644-1 or EU
GMP 2009.
2.
Press REPORT ENABLE to turn on green check (Figure 5-84).
3.
Press the REPORT button, which will display the saved reports
list.
Figure 5-85 Reports Screen
4.
Press ADD, which displays the Data View Buffer screen.
Summary of Contents for ApexP3
Page 1: ...ApexPortable P A R T I C L E C O U N T E R OPERATING MANUAL ApexP3 ApexP5...
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions ApexPortable Airborne Particle Counter Operating Manual...
Page 6: ......
Page 12: ...Lighthouse ApexPortable Operating Manual ii 248083440 1 Rev 5...
Page 16: ...Lighthouse ApexPortable Operating Manual 1 4 248083440 1 Rev 5...
Page 26: ...Lighthouse ApexPortable Operating Manual 3 6 248083440 1 Rev 5...
Page 30: ...Lighthouse ApexPortable Operating Manual 4 4 248083440 1 Rev 5...
Page 87: ...Operate the ApexPortable 248083440 1 Rev 5 5 57 Figure 5 95 Failed Report Sample...
Page 92: ...Operate the ApexPortable 248083440 1 Rev 5 5 62...
Page 120: ...Lighthouse ApexPortable Operating Manual A 18 248083440 1 Rev 5...
Page 122: ...Lighthouse ApexPortable Operating Manual B 2 248083440 1 Rev 5...
Page 128: ...Lighthouse ApexPortable Operating Manual I 6 248083440 1 Rev 5...
Page 129: ...00...