Check & Adjust
FlexLine, 241
5. Press
REC
to measure to the target point.
6. Either:
• Press
MORE
to measure another set to the same target point. The final
adjustment values will be the calculated average from all the measurements.
• Press
OK
to save the new adjustment data, or
• Press
ESC
to exit without saving the new adjustment data.
Messages
The following are important messages or warnings that may appear.
The old and new calculated values are displayed.
Messages
Description
V-angle not suitable
for adjustment !
The vertical angle deviates from the required horizontal /
line-of-sight, or in face II the vertical angle deviates by more
than 5° from the target point. Aim at the target point with
an accuracy of min. 5° or, when adjusting the tilt axis, 27°
above or beneath the horizontal plane. Confirmation of the
message required.
Results out of
tolerance. Previous
values retained !
Computed values out of tolerance. The previous values are
retained and measurements should be repeated.
Confirmation of the message required.
Hz-angle not suitable
for adjustment !
Horizontal angle in face II deviates by more than 5° from
the target point. Aim on the target point with an accuracy
of min. 5°. Confirmation of the message required.