Appendix B: Specifications
If you have any questions about the suitability of your glass slides, please consult a Leica Biosystems Technical Services
representative. For instructions on using the AutoLoader, see
“Chapter 6: Using the AutoLoader” on page 36
.
Scanner Control Workstation Software Specifications
Feature
Details
Operating system
Windows 7 64-bit
Scanning software
Console / Controller
Viewing software
Aperio ImageScope DX
System Requirements
You must use the workstation provided with the scanner and its calibrated monitor to scan. eSlide viewing also requires
using a workstation and two calibrated monitors provided by Leica Biosystems.
Environment Specifications
The Aperio AT2 DX is designed to be operated under the following environmental conditions: indoor use only, MAINS supply
voltage fluctuations of up to
±
10% of 100 - 120 V and 220 - 240 V, and transient overvoltages (category II of IEC 60364-4-
443) typically present on the MAINS supply. Also refer to the table below:
Feature
Details
Transportation/storage temperature
-22 degrees Fahrenheit to 113 degrees Fahrenheit
-30 degrees Celsius to 45 degrees Celsius
Operating temperature
65 – 81 degrees Fahrenheit
18 – 27 degrees Celsius
System heat dissipation
Typical BTU/hour: 1968
Power use
Typical watts/hour: 580
Transportation/storage humidity
30% – 90%, non-condensed
Operating humidity
30% – 90%, non-condensed, for a standard working pathology laboratory environment.
Maximum elevation
Up to 2000 Meters (6561.7 feet)
Routine setup and alignment may be required by a Leica Biosystems Technical Services representative after shipping.