![Laversab 6600-NG User Manual Download Page 93](http://html1.mh-extra.com/html/laversab/6600-ng/6600-ng_user-manual_3905006093.webp)
6600-NG User’s Manual
125-9110A
Page 93
Command
Description
Data sent from
PC
Data sent from
6600-NG
Pm
Read the value for Pt offset m, where m is as
defined in the table below.
m
Pt offset
1
10% offset
2
20% offset
3
30% offset
4
40% offset
5
50% offset
6
60% offset
7
70% offset
8
80% offset
9
90% offset
None
Pt offset value
Pm=n
Set the Pt offset m to a value of n. This command
requires the calibration password to be set.
n
0 if OK; or
Error code
Dm
Read the value for Ps2 offset m, where m is as
defined in the table below.
m
Ps2 offset
1
10% offset
2
20% offset
3
30% offset
4
40% offset
5
50% offset
6
60% offset
7
70% offset
8
80% offset
9
90% offset
None
Pt offset value
Dm=n
Set the Ps2 offset m to a value of n. This
command requires the calibration password to be
set.
n
0 if OK; or
Error code