6600-NG User’s Manual
125-9110A
Page 48
Figure 3.34 : Pt Leak (2-min)
At the end of three minutes, the screen will appear similar to Figure 3.35 below.
Figure 3.35 : Pt Leak (3-min)
The parameters indicated in Figures 3.32 through 3.35 are explained below.
[1] Leak Timer: Shows the elapsed time since the Leak-check was started.
[2] Start value: The Actual value at the start of the leak-check (at time 0:00)
[3] Leak timer values: Currently set at 1, 2 and 3 minutes. Can be changed to values up to 20 minutes.
See Section 3.10.7 for details.
[4] The Actual value at the end of 1 minute
[5] Total leak in 1 minute
[6] The Actual value at the end of 2 minutes
[7] Total leak in 2 minutes
[8] The Actual value at the end of 3 minutes
[9] Total leak in 3 minutes
When
CANC
is pressed to exit out of Leak mode, the 6600-NG automatically goes back to Control mode.
Leak Time: 0.00
Control
Leak
Control
4
±
8
000