6600-NG User’s Manual
125-9110A
Page 39
Ps1 Rate target exceeds limit. Press
CLR
Note: When Ps1 is in pressure units, the commanded Ps1 Rate target is limited by the pressure-
equivalent of the Ps1 Rate target that has been set for altitude units (feet/min or meters/sec). This
ensures that the rate of change of Ps1 pressure does not exceed the target altitude rate.
Note: In pressure units, Ps1-Rate targets can be set between 0 and 30 inHg/min. (or equivalent)
If the Rate Target value needs to be changed in small increments, instead of entering a value, the arrow
keys may be used to Jog the Rate Target value up or down. If the current Ps1 Rate Target value is 2000
Feet/min, and the Lower screen appears as in Figure 3.13 above, pressing the
Up-Arrow
key will make
the screen appear as shown in Figure 3.14 below.
Figure 3.20 : Jog the Ps1 Rate Target value
In this screen, the Upper section has a new symbol (with Up & Down arrows) to the right of the Ps1 Rate
Target value field. This indicates that the Ps1 Rate Target value can be jogged up or down repeatedly, in
steps of 10 Feet/min, using the Arrow keys, until
CANC
is pressed. In this case, since the
Up-Arrow
key
was pressed, the Ps1 Rate Target value was incremented to 2010 Feet/min. Press
Up-Arrow
one more
time and the value will change to 2020. Press
Down-Arrow
two times and the value decrements to 2000.
Note: It is not possible to Jog a Rate Target value if it is blinking.
The incremental value for the Jog feature varies depending on Ps1 units and is shown in the table below.
Ps1-Rate Units
Jog increment
Feet/min
10
Meters/sec
0.1
29.9213
30.0000
Pt inHg
Pt
Ps2
Ps1
Measure
inHg/min
0.000
4.000
inHg
Feet
29.9213
45
Pt inHg
30.0000
1
inHg/min
Feet/min
0.000
0
4.000
2010
Measure
Measure
AC
92%
Hide/Show Ps2
000
Messages