102 / 123
Proline heating thermostat
9.2.4
Device data
Master
MNEmv
ShoWX
(
Section 7.5)
Various device datas are displayed.
Command
Device data
Settings
Device status
Device
data
Display
.
Text shows various actual temperatures in °C
from ext. Pt100 and the modules.
T cont. head and T heatsink are temperatures of
electronics in the Master in °C.
Pump power in Watts, pump speed in rpm, pump
current in ampere.
T
int
indicates the current internal bath temperature
in °C.
Mains voltage in % of nominal and mains frequen-
cy in hertz.
Level indicates the liquid level in the internal bath.
Low Voltage of power transformer, of the 5 V
supply and fan voltage in volt.
Cur. cons.: Mains current consumption in ampere.
9.2.5
Fault memory
For the analysis and localization of faults the Command version includes a fault memory in which up to 45 fault and
alarm messages are saved.
Command
Errorstore
Settings
Device status
Errorstore
Display
.
The last message is at the top.
Each message line can be marked with the cur-
sor keys. The message appears in plain text in
the footer.
Under “Source”, the CAN node is displayed
which signaled the fault.
Code is the number, which in the Master is
shown in the display until the cause has been
rectified.
Type: Alarm, Warning or Fault (Error).
T ext Pt
25.70 Tint
25,58
T ext analog
---.-- Mains U(%) 103.74
T ext serial
---.-- Mains Frequ.
50
T cont. head 39.80 Level
6
T heatsink
51.68 Low voltage 27.90
Pump Pow. 44.90 5V Supply
5.00
Pump rpm
5460 Fan Voltage
7.0
Pumpe Cur.
1.68 Cur. cons.
2,84
Pump Menu
End
T
set
T
fix
No Source Code Type Date
Time
10 Safety 2
Alarm ------------ -----------
9 Safety 4
Warn. 28.08.03 15:32:02
8 Contro. 32
Error 17.07.03 10:52:02
7 Contro. 3
Warn. 06.06.03 11:15:11
6 Contro. 9
Alarm 05.06.03 08:45:01
5 Contro. 3
Alarm 01.06.03 17:58:22
4 Contro. 4
Warn. 28.05.03 20:01:22
3 Contro. 5
Warn. 27.05.03 07:58:00
Low level
Pump Menu
End
T
set
T
fix
Summary of Contents for PROLINE P 5 C
Page 13: ...02 2022 Proline heating thermostat 13 123 Clear view thermostat PVL 24 Bridge thermostat PB C ...
Page 113: ...02 2022 Proline heating thermostat 113 123 Pump characteristics measured with water ...
Page 114: ...114 123 Proline heating thermostat 02 2022 Pump characteristics measured with water ...
Page 115: ...02 2022 Proline heating thermostat 115 123 12 Declaration of Conformity ...
Page 118: ......
Page 119: ......