80 / 123
Proline heating thermostat
Editing the control parameter sets
The change in the control parameters is explained in Section 7.13.1 / 7.13.2 (internal / external). Once the value has
been changed and confirmed, the set number, e.g.
Set 3
and
Upload actual
, the new value is accepted into the con-
trol parameter set to be changed (Set 3) via the command
Control parameter sets
.
7.14
Alarms, Warnings and Errors
The SelfCheck Assistant of your Proline Thermostat monitors more than 50 device parameters and triggers alarms,
warnings or errors as appropriate.
All warnings and alarms are shown on the Command remote control in plain text. Errors are shown in plain text on the
Command remote control in an error list.
Alarms:
Alarms are safety relevant. Pump and heater unit will be shut off.
Warnings:
Warnings normally are not safety relevant. The thermostat continues to operate.
Errors:
When an error occurs, switch of the device. If the error is always present after switching on the de-
vice, please inform the LAUDA Service (
Find cause of alarm or warning and rectify where necessary. Then press on the Master keyboard in order to remove
the alarm message. Warning messages can be removed either on the Master keyboard with or on the Command
board with .
Warnings may be ignored by pressing or on the Master keyboard or by activating the Screen Softkey on the
Command remote control. Otherwise warnings will be repeated periodically.
7.14.1
Overtemperature protection and checking
The units are designed for operation with non-flammable and flammable liquids to DIN EN 61010-2-
010.
Setting the overtemperature cut-off: Recommended setting: 5 °C above desired
bath temperature.
Caution! The overtemperature switch point T
max
is being controlled by a system
that works independent of the internal bath control. The setpoint setting can be
limited independently to T
max
with the functions T
ih
and T
il
(
The cut-off point is displayed in the LED display on pressing the key
.
Changing the overtemperature cut-off point:
For safety, and to guard against unintentional adjustment, the key
must
be held pressed during all the following entries. Now, briefly press . The
display flashes and the overtemperature cut-off can be set with the keys
or .
Overtemperature
cut-off point
90
°C
Summary of Contents for PROLINE P 5 C
Page 13: ...02 2022 Proline heating thermostat 13 123 Clear view thermostat PVL 24 Bridge thermostat PB C ...
Page 113: ...02 2022 Proline heating thermostat 113 123 Pump characteristics measured with water ...
Page 114: ...114 123 Proline heating thermostat 02 2022 Pump characteristics measured with water ...
Page 115: ...02 2022 Proline heating thermostat 115 123 12 Declaration of Conformity ...
Page 118: ......
Page 119: ......