56 / 123
Proline heating thermostat
Confirm the control dialog on the right with
or cancel with .
Return to the measurement window with
End or .
7.8.9
Entering the offset of the external temperature probe
If a deviation is found during the check using a calibrated reference thermometer probe, then the offset (the additive
part of the characteristic) of the external measurement chain can be adjusted with the following function. The refer-
ence thermometer must be dipped nearly by the external temperature probe into the consumer bath according to the
details on the calibration certificate.
Master
CAL E
3x
to
MNEnu
2x
to
CAL..
2x
Call the offset adjustment for the external temperatures probe
CAL E.
Continue as described in (
7.8.7) for the internal temperature probe.
Command
Calibration
Open the device parameter menu via the soft key
Menu .
With the cursor keys continue to:
Settings
Calibration
extern Pt100
Calibration
.
The adjacent window appears.
Confirm selection with .
Continue as described in (
7.8.7) for the inter-
nal temperature probe.
Pump Menu
End
T
set
T
fix
Confirm input!
Enter key: Continue
Escape key: Cancel
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Summary of Contents for PROLINE P 5 C
Page 13: ...02 2022 Proline heating thermostat 13 123 Clear view thermostat PVL 24 Bridge thermostat PB C ...
Page 113: ...02 2022 Proline heating thermostat 113 123 Pump characteristics measured with water ...
Page 114: ...114 123 Proline heating thermostat 02 2022 Pump characteristics measured with water ...
Page 115: ...02 2022 Proline heating thermostat 115 123 12 Declaration of Conformity ...
Page 118: ......
Page 119: ......