Proline heating thermostat
67 / 123
If the field in the column
Time
has a black
background, the entry mode for the "Segment
time" time setting is obtained by pressing the
key.
If 0 is entered into the field "Time", ------ ap-
pears. Then the final temperature is ap-
proached as quickly as possible. With a time
entry the final temperature is obtained exactly
after the time expires (ramp).
Enter the segment time and confirm with the
key.
Continue to the "Tolerance" entry field with .
If the field in the column "Tolerance" has a
black background, the entry mode for the
"Temperature tolerance" is obtained by press-
ing the key. It defines how accurately the
end of segment temperature is to be obtained
before the next segment is processed.
A tolerance which is selected too small can
stop the next segment from being started ac-
cording to plan.
Set the temperature tolerance and confirm
with .
Continue with to the entry field "Pump".
If the field in the column "Pump" has a black
background, the entry mode for the
Pump lev-
el
is obtained by pressing the key .
With or select Pump Level 1 – 8 or ---
---- and confirm with .
------ stands for "no change to previous seg-
ment", i.e. when ------ is present in all fields,
the pump level always retains the start setting
or the setting before the program start.
Continue with to the field "Out 1", "Out 2"
or "Out 3".
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Temp. tolerance (0=off):
10,
00
Min: 0,00°C Max:450,00°C
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Input segment time:
003:00
Hours(max.999):Minutes
Pump Menu
End
T
set
T
fix
Level 8
Level 7
Level 6
Level 5
Level 4
Level 3
Level 2
Level 1
-------
Pump level
Summary of Contents for PROLINE P 5 C
Page 13: ...02 2022 Proline heating thermostat 13 123 Clear view thermostat PVL 24 Bridge thermostat PB C ...
Page 113: ...02 2022 Proline heating thermostat 113 123 Pump characteristics measured with water ...
Page 114: ...114 123 Proline heating thermostat 02 2022 Pump characteristics measured with water ...
Page 115: ...02 2022 Proline heating thermostat 115 123 12 Declaration of Conformity ...
Page 118: ......