Quit the change mode by pressing again or automatically after 5 seconds,
while you keep
pressed.
This somewhat complicated procedure is intended to prevent unintentional
adjustment.
Set the overtemperature switch-off point Tmax below the flash point of the heat transfer liquid
(
The setting range is restricted to 5 °C above the upper limit of the working temperature range (Tih
If the bath temperature rises above the overtemperature cut-off:
1.
Alarm sounds as dual-tone signal.
2.
tEMNP
for overtemperature appears in the display.
3.
Red LED
above the fault triangle flashes.
Heater switches off on both poles,
Pump is switched off electronically.
Rectify cause of fault.
Wait until the bath temperature has cooled below the cut-off point or set the
cut-off point higher than the bath temperature. When
tEMNP
is shown
in the display:
Unlock with the key.
Unlocking is not possible on the Command remote control!
Before longer periods of unsupervised operation, the overtemperature protection should be checked.
To do this:
Slowly lower T
max
, as described above.
Cut-off at the bath temperature should occur.
Step 1 – 2 (see above) must follow.
Set the overtemperature cut-off higher than the bath temperature again and wait
until
tEMNP
appears in the display.
Unlock with the key.
Unlocking is not possible on the Command remote control!
Command
Overtemperature alarm!
Overtemperature alarm! is shown in the display and signifies that unlocking is only
possible on the Master control panel.
Overtemperature alarm
tEMNP
°C
Summary of Contents for PROLINE P 5 C
Page 13: ...02 2022 Proline heating thermostat 13 123 Clear view thermostat PVL 24 Bridge thermostat PB C ...
Page 113: ...02 2022 Proline heating thermostat 113 123 Pump characteristics measured with water ...
Page 114: ...114 123 Proline heating thermostat 02 2022 Pump characteristics measured with water ...
Page 115: ...02 2022 Proline heating thermostat 115 123 12 Declaration of Conformity ...
Page 118: ......
Page 119: ......