Proline heating thermostat
7.8.7
Entering the offset of the internal temperature probe
If, during checking with a calibrated reference thermometer probe, a deviation is found, then the offset (i.e. the additive
part of the characteristic) of the internal measuring chain can be adjusted with the following function. The reference
thermometer must be dipped into the bath according to the details on the calibration certificate.
Master
CAL 1
3x
to
MNEnu
2x
to
CAL..
Call the offset adjustment of the internal temperature probe
CAL..
.
CAL 1
is displayed.
3 sec
The actual value of the bath temperature is displayed flashing.
or
Set the value which you have read from the calibrated reference measure-
ment device (with glass thermometers consider the correction where applica-
ble!).
The entered value is accepted and
donE
is displayed.
Command
Calibration
Open the device parameter menu via the soft
key Menu .
With the cursor keys continue to:
Settings
Calibration
intern Pt100
Calibration
.
The window shown on the left appears.
Confirm selection with .
CAL 1
°C
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Summary of Contents for PROLINE P 5 C
Page 13: ...02 2022 Proline heating thermostat 13 123 Clear view thermostat PVL 24 Bridge thermostat PB C ...
Page 113: ...02 2022 Proline heating thermostat 113 123 Pump characteristics measured with water ...
Page 114: ...114 123 Proline heating thermostat 02 2022 Pump characteristics measured with water ...
Page 115: ...02 2022 Proline heating thermostat 115 123 12 Declaration of Conformity ...
Page 118: ......
Page 119: ......