n
[Temperature scale]: automatic or manual.
Here you can define which temperature range is displayed in the visible
graph window.
[Autom.]: The size of the visible graphic area automatically adapts to
the changing temperature curves.
If the setting is selected automatically, the following menu item
(Temperature limits) is not visible.
n
[Temperatutre limits]: T.scale Min and T.scale Max.
Here you can manually define which time range is displayed in the visible
graph window.
5.3.4 Operate the device using the Command Touch unit
The device can also be operated with the Command Touch remote control
unit (accessory).
WARNING!
Device start carried out from remote control unit
Scalding, cold burns
Switch the local operating unit into standby
5.4 Setting heat transfer liquid
WARNING!
Overheating of heat transfer liquid due to incorrect entry of
overtemperature switch-off point T
max
for the hydraulic circuit
Fire
Set the overtemperature switch-off point for the hydraulic
circuit to 5 K above the upper limit of the temperature
range for your application. But not above the upper limit of
the working temperature range of the heat transfer liquid.
The following information is only relevant for the Integral T device:
WARNING!
Overheating of heat transfer liquid due to incorrect entry of
overtemperature switch-off point T
max
for the hydraulic circuit
Fire
In the menu of the device, set the heat transfer liquid used
in the device. Set the overtemperature switch-off point
below the flash point of the heat transfer liquid.
The following information is only relevant for the Integral XT device:
V6
Integral Process Thermostats and High-Temperature Thermostats
69 / 198