ID Function
Unit
Command
119 Solenoid valve for automatic filling device
(Solenoid valve must be present)
[–]
VERSION_M_1
120 Solenoid valve for constant level device
(Solenoid valve must be present)
[–]
VERSION_M_2
121 Solenoid valve, shut off valve 1
(Solenoid valve must be present)
[–]
VERSION_M_3
122 Solenoid valve, shut off valve 2
(Solenoid valve must be present)
[–]
VERSION_M_4
124 Pump 1
[–]
VERSION_P_0
125 Pump 2
[–]
VERSION_P_1
126 Heating system 1
[–]
VERSION_H_0
127 Heating system 2
[–]
VERSION_H_1
128 External Pt100 interface 1
(module must be present)
[–]
VERSION_E
129 External Pt100 interface 2
(second module must be present)
[–]
VERSION_E_1
4.7.7 Write commands
The interface module recognizes the following write commands, which you
can use to transfer values to the constant temperature equipment.
The control station obtains operator privileges when sending a write
command. This only works if the control station has not been locked
by another control element
Chapter 6.20 “Operator and viewer”
The constant temperature equipment confirms each write command
with OK, e.g. the reply from the device address A015 is
"A015_OK". In the event of an error, an error message is output
as a reply, e.g. "A015_ERR_6", see
.
Table 23: Temperature
ID Function
Unit
Command
1 Temperature set point
[°C]
OUT_SP_00_XXX.XX
15 Actual value of external temperature (via interface)
[°C]
OUT_PV_05_XXX.XX
26 Limitation of outflow temperature TiH (upper limit)
[°C]
OUT_SP_04_XXX
V6
Integral Process Thermostats and High-Temperature Thermostats
58 / 198