Please note:
n
If in the
hh
and
:mm
field the value "0" is entered, the temperature T
end
will be started as quickly as possible.
n
Changes to the pump level are entered in the respective segment (=
program line).
n
The default value of the contact module is
off
.
You have the following options in the selected program:
n
By pressing the right arrow key 5 times you can display the columns
Pump, S1, S2 and S3 of the program.
n
Use the left arrow key to display the columns Tend, hh, :mm and Toler-
ance again.
n
With the [up] and [down] arrow keys, you can navigate to the segments
(lines) of a program.
n
With the [OK] you select a value for editing.
n
Use the [right] and [left] arrow keys to select individual digits of the
value.
n
With the [up] and [down] arrow keys, you can increase or reduce the
selected digit.
n
With the [ESC] softkey, you can deselect a selected value again.
n
With the [OK] key, you confirm your change.
n
You exit the program with the [ESC] softkey. The entered values are
saved.
1.
Navigate to the segment under which the new segment should be
added.
2.
In this segment, navigate to the column with the
No.
3.
Press the [NEW] key.
A new segment is created.
1.
Navigate to the segment that you want to delete.
2.
In this segment, navigate to the column with the
No.
3.
Press the
DELETE
softkey.
The segment is deleted.
Editing a program
Add new segment
Fig. 66: Select program segment
Delete segment
V6
Integral Process Thermostats and High-Temperature Thermostats
113 / 198