44 / 124
ECO Silver
Caution!
Bath overflow due to thermal expansion or
immersion of objects
Burns, scalds, frostbite
Take the volume of external consumers into account.
Take into account the increase in volume with a rise in tem-
perature.
Caution!
Hot vapor formation / discharge of boiling cooling-water on the cool-
ing coil
Burns, scalds
Filling of cooling coil with cooling water only admissible up to
T
max
of 100 °C!
Notice
Inadmissible operating temperatures;
temperature difference between outflow and product
too large
Property damage (consumers, external components)
Note that an externally controlled bath temperature, espe-
cially during a transient response, may differ substantially
from the set-point temperature.
Note the various limitation options (Tih, Til, T
max
, correction
limitation).
Set the overtemperature switch-off point T
max
according to
the heat transfer liquid. T
max
must be below the flash point.