ECO Silver
95 / 124
Creating and editing a program:
Compare the programming example (
–
You access the editor view of the programmer by selecting and
confirming
Programmer
Edit
. To view the complete win-
dow information go to the right with .
–
With the keys , , and you obtain access to the individual
segment fields.
–
If the cursor is located in the first column, the functions "new" and
"delete" are visible in the soft-key bar. Program steps can be creat-
ed or deleted in this way.
–
With all other cursor positions the function "OK" is visible in the
soft-key bar.
–
Using (OK) you can select the appropriate parameters and
change them with or .
–
You can quit the edit window at any time without changes using (ESC). When the cursor is located on a segment
number, using you return to the menu level of the programmer without changes.
Note: No time specification is possible in the start segment. The temperature of the first segment is attained as quickly as
possible in order to switch to segment 2 after reaching the set tolerance.
The programmer edit window contains the following parameters:
The segment number of the program is located in the first column.
Tend:
Final temperature to be attained
hh:mm:
Time in hours (hh) and minutes (mm) in which the specified temperature is to be attained.
If the value "0:00" is entered in the field "hh:mm", the set value is accepted immediately
and the bath temperature approached as quickly as possible.
Tol.:
Defines how exactly the final temperature is to be attained before the next segment it processed.
If the tolerance range is selected too small in the "Tol." field, the program might not continue, because the
required tolerance is not achieved.
Pmp:
Pump level at which the segment is to be processed.
S1, S2, S3: Switching contacts of the contact module (if present) can be programmed here. Contact modules are availa-
ble as accessories (
9). The setting "- -" stands for no change to the preceding segment, i.e. if "- -" is pre-
sent in all fields, the contact setting of the start setup or that before the program start is retained.
–
Inserting a new segment
–
–
With or go to the segment number under which the new seg-
ment is to be inserted.
–
The new segment is inserted on pressing (new). You can edit it as
described above.
–
Deleting a segment
Tend
hh:mm Tol.
1
35.00
--:--
0.1
2 40.00
00:10
0.1
3 50.00
00:20
0.0
4 70.00
00:50
0.1
ESC
NEW
DELETE
Pmp
S1
S2
S3
1
1
on
- - -
off
2
1
on
- - -
off
3
2
on
- - -
off
4
2
on
- - -
off
ESC
OK
- - -
Tend
hh:mm Tol.
1
35.00
--:--
0.1
2 40.00
00:10
0.1
3 50.00
00:20
0.0
4 70.00
00:50
0.1
ESC
new
DELETE
Summary of Contents for ECO Silver E 10 S
Page 1: ...ECO Silver Operation manual ...
Page 2: ......
Page 4: ......
Page 78: ...78 124 ECO Silver 10 2019 i ...
Page 79: ...10 2019 i ECO Silver 79 124 ...
Page 80: ...80 124 ECO Silver 10 2019 i Appendix with settings ...
Page 111: ...10 2019 i ECO Silver 111 124 5 Click on Continue installation 6 Click on the key Finish ...
Page 113: ...10 2019 i ECO Silver 113 124 ...
Page 120: ...120 124 ECO Silver 10 2019 i Blank page ...
Page 121: ...10 2019 i ECO Silver 121 124 Blank page ...
Page 122: ...122 124 ECO Silver 10 2019 i ...
Page 125: ......
Page 126: ......
Page 127: ......