114 / 124
ECO Silver
Command
Meaning
OUT_PV_05_XXX.XX
Specify external temperature via interface
OUT_SP_00_XXX.XX
Set-value transfer with max. 3 places before the decimal point and max. 2 places after it.
OUT_SP_01_XXX
Pump power level 1 to 6
OUT_SP_02_XXX
Cooling operating mode (0 = OFF / 1 = ON / 2 = AUTOMATIC).
OUT_SP_04_XXX
TiH outflow temperature limit, upper value
OUT_SP_05_XXX
TiL outflow temperature limit, lower value
OUT_PAR_00_XXX.X
Setting of the control parameter Xp.
OUT_PAR_01_XXX
Setting of the control parameter Tn (5 – 180 s; 181 = Off).
OUT_PAR_02_XXX
Setting of the control parameter Tv.
OUT_PAR_03_XX.X
Setting of the control parameter Td.
OUT_PAR_04_XX.XX
Setting of the control parameter KpE.
OUT_PAR_05_XXXX
Setting of the control parameter TnE (0 – 9000 s; 9001 = Off).
OUT_PAR_06_XXXX
Setting of the control parameter TvE (5 = OFF).
OUT_PAR_07_XXXX.X
Setting of the control parameter TdE
OUT_PAR_09_XXX.X
Setting of the max. correcting quantity limit.
OUT_PAR_10_XX.X
Setting of the control parameter XpF.
OUT_PAR_14_XXX.X
Setting of the setpoint offset.
OUT_PAR_15_XXX
Setting of the control parameter PropE.
OUT_MODE_00_X
Keypad: 0 = released / 1 = locked (corresponds to: "KEY").
OUT_MODE_01_X
Control: 0 = int. / 1 = ext. Pt100 / 2 = ext. Analog / 3 = ext. serial.
OUT_MODE_03_X
Keypad Command remote control: 0 = released / 1 = locked.
OUT_MODE_04_X
Setpoint offset source: 0 = normal / 1 = ext. Pt / 2 = ext. analog / 3 = ext. serial.
START
Switches the device on (from Standby)
STOP
Switches the device in Standby (pump, heating, chiller off).
RMP_SELECT_X
Selection of program (1 — 5) to which further commands are to refer. When the device is
switched on Program 5 is selected.
RMP_START
Start the programmer.
RMP_PAUSE
Stop the programmer.
RMP_CONT
Start the programmer again after a hold.
RMP_STOP
Terminate the program.
RMP_RESET
Delete program (all segments)
RMP_OUT_00_XXX.XX_XXXXX_XXX.XX_X
Sets programmer segment (temperature, time, tolerance, and pump level). A segment is
appended and assigned appropriate values.
RMP_OUT_02_XXX
Number of program loops: 0 = endless / 1 — 250.
F.
7.
1
Interface write commands (data issued to the thermostat)
F
.
7
Commands and error messages applicable to the RS 232/485 interface module and to the
Ethernet interface
Summary of Contents for ECO Silver E 10 S
Page 1: ...ECO Silver Operation manual ...
Page 2: ......
Page 4: ......
Page 78: ...78 124 ECO Silver 10 2019 i ...
Page 79: ...10 2019 i ECO Silver 79 124 ...
Page 80: ...80 124 ECO Silver 10 2019 i Appendix with settings ...
Page 111: ...10 2019 i ECO Silver 111 124 5 Click on Continue installation 6 Click on the key Finish ...
Page 113: ...10 2019 i ECO Silver 113 124 ...
Page 120: ...120 124 ECO Silver 10 2019 i Blank page ...
Page 121: ...10 2019 i ECO Silver 121 124 Blank page ...
Page 122: ...122 124 ECO Silver 10 2019 i ...
Page 125: ......
Page 126: ......
Page 127: ......