84 / 124
ECO Silver
–
Access to the main menu level is obtained by pressing the enter
key .
–
Selection and confirmation of
Setup
Basic setup
Curr.Consumpt.
.
The entry window appears.
–
Change the value with or .
–
Single figures can be selected by pressing or .
–
Confirm the input with the enter key .
–
By pressing (ESC) you are returned to the menu level without any change.
Deviations to the calibrated reference thermometers (e.g. LAUDA DigiCal) can be corrected internally by the "Off-
set" function.
–
You activate the soft-key bar by pressing any key.
–
Access to the main menu level is obtained by pressing the enter
key .
–
Selection and confirmation of
Setup
Calibration
.
The adjacent menu window appears.
–
Select
Calibration
with or and confirm with .
The entry window appears. The value indicated on the reference ther-
mometer must be entered as the value.
–
Change the value with or .
–
Single figures can be selected by pressing or .
–
By pressing (+/-) the arithmetic sign can be changed.
–
You confirm the set value by pressing .
–
By pressing (ESC) you are returned to the menu level without any change.
Curr.Cons.
Min: 8.00
Max: 16.00
ESC
OK
- - -
16.00
Temp. ref. device
Min: -22.02Max: 28.02
ESC
OK
+ / -
25.02
ESC
OK
STOP
Calibration
Factory calibration
A
.
7
Entering the offset of the displayed temperature (calibration)
Summary of Contents for ECO Silver E 10 S
Page 1: ...ECO Silver Operation manual ...
Page 2: ......
Page 4: ......
Page 78: ...78 124 ECO Silver 10 2019 i ...
Page 79: ...10 2019 i ECO Silver 79 124 ...
Page 80: ...80 124 ECO Silver 10 2019 i Appendix with settings ...
Page 111: ...10 2019 i ECO Silver 111 124 5 Click on Continue installation 6 Click on the key Finish ...
Page 113: ...10 2019 i ECO Silver 113 124 ...
Page 120: ...120 124 ECO Silver 10 2019 i Blank page ...
Page 121: ...10 2019 i ECO Silver 121 124 Blank page ...
Page 122: ...122 124 ECO Silver 10 2019 i ...
Page 125: ......
Page 126: ......
Page 127: ......