6-10
Data Storage After Improper Shutdown
LxT Manual
Improper Shutdown
Following an improper shutdown, when the instrument is
next turned On:
Step 1
The user is prompted to save the data.
• If the user responds by selecting to store the data, the
data is stored and the instrument is reset
• If there is no user response to the prompt within ten
seconds, the instrument is reset
• If the user responds by selecting not to store the data, the
sequence moves to Step 2.
Step 2
The user is prompted to reset the instrument
• If the user responds by selecting to reset the instrument,
the instrument is reset.
• If there is no user response to the prompt within ten
seconds, the instrument is reset.
When present, the user must
eventually select to store the data,
reset the instrument, or take no
action, in which case the instrument
is automatically reset.
• If the user responds by selecting not to reset the
instrument, the sequence moves back to Step 1.
This sequence is diagramed below
Summary of Contents for SoundTrack LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...