LxT Manual
Technical Specifications
A
-
11
Min/Max Integration Time
Minimum and maximum integration time for measurement of time-average levels and sound exposure
levels.
DC Output Voltage Range
0 to +3 Volt (0 to 300 dB)
V0 = SPL/100
SPL = 100*V0
or
Sensitivity = 0.01 V/dB with resolution of 0.001 V
DC Output Impedance
3650
DC Output Recommended Loads
for less than 0.4% error
A 10
DC Voltmeter represents negligible error (-0.036%).
The output resistance can be accounted for in the interface
design. For example, if a chart recorder has an input load of
10 k
the gain can be set to 1.365 to correct for the loading.
Time Average Levels and Sound Exposure Levels, (s)
Minimum
0.1
Maximum
(daily autostore enabled)
Unlimited
Maximum
(daily autostore disabled, errors less
than 0.5 dB)
> 23 days
Dosimeter Metrics: TWA, Dose (s)
Minimum
0.1
Maximum
Unlimited
Table A-10 Min/Max Integration Time
Table A-9 AC/DC Output
Summary of Contents for SoundTrack LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...