7-8
Acoustic Calibration
LxT Manual
Set the CAL200 level switch to 94 or 114 dB.
Adding a Calibrator
When adding a calibrator to the list, the following
information may be entered:
• Calibration Level
• Calibration Frequency
• Calibrator Description
The calibration level and frequency values are as specified
in the section ‘Recommended Calibrators for Use with LxT1
and LxT2” on page 7-5.
On the
Calibrate
tab, select each text box in the
Calibrator
section and enter the correct information about a calibrator;
press
5
to complete the entry, as shown in Figure 7-3.
When the calibration level, calibration frequency, and
calibrator description have been entered, select the
Save
button and press
5
to save the information to the list of
calibrators.
Class 2, LxT2 with 375B02
microphone
Static Pressure
65 kPa to 108 kPa
9.4 psi to 15.7 psi
Temperature
0
C to + 40
C
32
F to + 104
F
Relative Humidity
25% to 90%, without condensation
from - 10
C to + 39
C
(14
F to + 102
F)
Instrument Class
Parameter
Range
Table 7-2 Environmental Parameter Ranges for Calibration
Summary of Contents for SoundTrack LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...